18

Correct posture

• Rest for approx. 5 minutes before each measurement. 

Otherwise there may be divergences. 

•  You can perform the measurement 

while sitting or lying. To carry out a 

blood pressure measurement, make 

sure you are sitting comfortably with 

your arms and back leaning on 

something. Do not cross your legs. 

Place your feet flat on the ground. Make sure to rest 

your arm and move it. Always make sure that the cuff is 

at heart level. Otherwise significant deviations can 

occur. Relax your arm and the palm of your hand.

• In order not to distort the result, it is important to keep 

still during the measurement and not talk.

Measuring blood pressure 

• Switch on the blood pressure monitor with the   but-

ton.

•  Before the measurement, the last saved test result is 

briefly displayed. If there is no measurement in the mem-

ory, the instrument always displays the value  .

• The cuff automatically inflates. Cuff air pressure is 

released slowly. If a tendency towards high blood pres-

sure is already detectable, the cuff is pumped up again 

and cuff pressure increased further. As soon as a heart 

rate is detected, the heart rate symbol   is displayed. 

• Heart rate, systolic and diastolic blood pressure are 

displayed. 

• You can interrupt measurement at any time by press-

ing the   button.

•  _ appears if it has not been possible to perform the 

measurement properly. Observe the section in these 

instructions on error messages/troubleshooting and 

repeat the measurement.

 

• The test result is saved automatically.

Evaluating results
Cardiac arrhythmia:

This instrument can identify possible cardiac arrhythmia 

disorders during measurement and if necessary indicates 

the measurement with the flashing icon 

.

 

This may be an indicator for arrhythmia. Arrhythmia 

is a condition where the heart rhythm is abnormal as 

a result of defects in the bioelectrical system control-

ling the heart beat. The symptoms (omitted or prema-

ture heart beats, slow or excessively fast heart rate) may 

be caused, among other things, by heart disease, age, 

physical predisposition, excessive use of stimulants, 

stress or lack of sleep. Arrhythmia can only be ascer-

tained through examination by your doctor. 

Repeat the measurement if the flashing icon 

 is 

displayed after the measurement. Please note that you 

should rest for 5 minutes between measurements and 

• • • • • • • • • • • • •

 

 

 

Содержание SBM 03

Страница 1: ...33 I Misuratore di pressione Istruzioni per l uso 34 43 r Прибор для измерения кровяного давления на запястье Инструкция по применению 44 55 Electromagnetic Compatibility Information 56 58 Serviceadresse Hans Dinslage GmbH Riedlinger Straße 28 88524 Uttenweiler GERMANY Tel Nr 0 7374 915766 E Mail service sanitas online de SBM 03 ...

Страница 2: ...e suffering from cardiac arrhythmia The values determined are classified and graphically evaluated according to WHO guidelines Keep these instructions carefully for further use and also let other users have access to them 2 Important information Signs and symbols The following symbols are used in these instructions for use on the packaging and on the type plate for the device and accessories Cauti...

Страница 3: ...edicines and their administration Do not use the blood pressure monitor on newborns or patients with preeclampsia We recommend consulting a doctor before using the blood pressure monitor dur ing pregnancy In the case of restricted circulation on the arm as a result of chronic or acute vascular diseases including vascular constriction the accuracy of the wrist meas urement is limited In this case y...

Страница 4: ...nstrument s service life depend on careful hand ling You should protect the device from impact moisture dirt major temperature fluctuations and direct expo sure to the sun s rays Never drop the device Do not use near strong electromagnetic fields i e keep it away from any radio systems and mobile phones Only ever use the cuffs provided with the monitor or original replacement cuffs Otherwise erron...

Страница 5: ...em if necessary prior to making any complaint The appliance should be disposed of accor ding to Regulation WEEE Waste Electrical and Electronic Equipment In case of queries please contact the municipal authorities responsible for waste disposal in your area 3 Unit description 1 Display 2 Memory button M 3 START STOP 4 Battery cover 5 Wrist cuff Icons in the display 1 Time and date 2 Systolic press...

Страница 6: ...his way is it pos sible to save and subsequently retrieve your measured values with the right date and time Time is shown in 24 hour format Note If you hold down M you can set the values faster Press and M simultaneously The month starts flashing Using the M button set the month 1 to 12 and confirm with Set the day hour and minute and confirm in each case with 5 Measuring blood pressure Positionin...

Страница 7: ...eased further As soon as a heart rate is detected the heart rate symbol is displayed Heart rate systolic and diastolic blood pressure are displayed You can interrupt measurement at any time by press ing the button _ appears if it has not been possible to perform the measurement properly Observe the section in these instructions on error messages troubleshooting and repeat the measurement The test ...

Страница 8: ...to WHO Guidelines Definitions and the latest findings the test results can be classified and evaluated according to the following chart The bar graph in the display and the scale on the unit indicate the range of the blood pressure which has been recorded An accoustic evaluation of your measurement also follows If the values for systolic and diastolic pressure are in two different WHO ranges e g s...

Страница 9: ...tightly or loosely or appears on the display the pump pressure is higher than 300 mmHg appears on the display pumping up takes longer than 160 seconds appears on the display there is a system or device error or appears on the display the batteries are almost empty In the above cases you must repeat the measurement Reinsert the batteries if necessary or else replace them Technical alarm description...

Страница 10: ...ssible operating conditions 5 C to 40 C 90 relative air humidity non condensing Permissible storage conditions 20 C to 55 C 95 rela tive air humidity 800 1050 hPa ambient pressure Power supply 2x 1 5 V AAA batteries Battery life For approx 180 measurements depending on the blood pressure level and or pump pressure Accessories Instruction for use 2x 1 5 V AAA batteries storage box Classification In...

Страница 11: ...pressure measur ing systems and IEC 80601 2 30 Medical electrical equipment Part 2 30 Particular requirements for the safety and essential performance of automated non invasive blood pressure monitors The accuracy of this blood pressure monitor has been carefully checked and developed with regard to a long useful life If using the device for commercial medical purposes it must be regularly tested ...

Отзывы: