VIII. Simple Troubleshooting

Failure

Failure Cause

Approach

Machine turned on but

unelectrified

Power cord is not plugged

in place

Check and ensure the power interface

and the power plug is plugged in

place and in good contact.

Fuse burnt-out

Open the electric box cover, check the

fuse and replace with a spare one if

necessary. Check whether the power

supply voltage is stable; Check and

ensure the power interface and the

power plug are in good contact.

Flow Alarm. Use a water pipe

to directly connect to the

water outlet and inlet but still

without water flowing

Water level in the water

tank is too low

Check the water level gauge display,

add water until the level shows in the

green area; And check whether water

circulation pipe leaks.

Flow alarm occurs while

chiller is connecting toother

equipment, but there is water

flowing and no alarm when a

water pipe is directly

connected to the chiller

water outlet and inlet.

Water circulation pipes are

blocked or a pipe bending

deformation.

Check water circulation pipe

Ultrahigh water temperature

alarm

Blocked dust gauze, bad

thermolysis

Disassemble and clean the dust gauze

regularly

Poor ventilation for air

outlet and inlet

To ensure a good ventilation

for air outlet and inlet

Voltage is extremely low or

unstable

To improve the power supply circuit or

use a voltage regulator

Improper parameter

settings on thermostat

To reset controlling parameters or

restore factory settings

Switch the power on and

off frequently

To ensure there is sufficient time

for refrigeration

(more than 5 minutes)

Excessive heat load

Reduce the heat load or use other

model with larger cooling capacity

Ultrahigh room temperature

alarm

The room temperature is

too high for the chiller

To improve the ventilation to

guarantee that the machine is running

below 40

room temperature.

Serious problem of

condensate water

Water temperature is much

lower than ambient

temperature, high humidity

Increase water temperature or

to preserve heat for pipeline

Water drains slowly from

drain port during water

changing

Water supply inlet is not

open

Open the water supply inlet

Startup flow alarm

Bubble in the water pipe

Exhaust the air

Содержание CW-5300

Страница 1: ...TY JS 20319 003 1 CW 5300 6000 6100 6200 6260 Industrial Chiller User Manual...

Страница 2: ...rly so that you can refer to it whenever you need information This manual is not a quality guarantee Our company reserves the right to the interpretation of the correction of misprint and improperly d...

Страница 3: ...4 I Overview 7 II Model Illustration 7 III Product Outlines Parts 8 IV Installation 9 V Operating States Parameters Adjustment 11 VI Alarm Output Terminal 14 VII Maintenance 15 VIII Simple Troublesho...

Страница 4: ...on a The power supply must conform to the standard indicated in the nameplate or the manual b Use the standard cable and select the standard wire diameter c The grounding wire must be installed and th...

Страница 5: ...ilable Otherwise it may cause anesthesia and suffocation if a large amount of refrigerant fills the enclosed space b Avoid contacting with skin or it will cause frostbite DANGER Failure to take safety...

Страница 6: ...r filter at least once a week If it is blocked the cooling ability will decrease and the power consumption will increase and the alarm will not work properly Equipment surface cleaning a Please use cl...

Страница 7: ...hiller the refrigerant in the evaporator coil absorbs the heat of the return water and vaporizes into steam The compressor continuously extracts the generated steam from the evaporator and compresses...

Страница 8: ...mes are only for one of the models The outlines and parts installation may be slightly different in different models Power Switch Temperature Controller Air Inlet Filter Gauze Water Supply Inlet Elect...

Страница 9: ...etc 4 Operating Environment Ambient Temperature Ambient Humidity Altitude 0 40 90 RH 3 000m 5 Medium Requirements Cooling medium allowed by the chiller includes purified water distilled water high pu...

Страница 10: ...1 Please select the cable based on the max rated current indicated on the label of the chiller 2 Recommended reference standard for power cord diameter selection Rated current A 5 10 15 25 35 50 Wire...

Страница 11: ...control mode It will self adjust controlling parameters according to room temperature for meeting equipment cooling requirements The new T 506E intelligent temperature controller is designed with con...

Страница 12: ...ings After the power is on for 30 seconds press and hold and keys for 3 seconds until it displays rE The supplier parameter and user parameter will restore to factory value 3 seconds later it will ret...

Страница 13: ...perature is 5 higher than the set water temperature and the alarm of over low water temperature will be on when water temperature is 10 lower than the set water temperature Users can easily adjust wat...

Страница 14: ...rly disassemble the filter gauze for cleaning and use a compressed air gun to blow away the dust on the condenser about 1 week is suggested and the air filter must not be missing for a long time Displ...

Страница 15: ...ending deformation Check water circulation pipe Ultrahigh water temperature alarm Blocked dust gauze bad thermolysis Disassemble and clean the dust gauze regularly Poor ventilation for air outlet and...

Страница 16: ...TY JS 20319 003 16 Attachment Schematic CW 5300 flow sensor CW 5300 flow switch...

Страница 17: ...TY JS 20319 003 17 CW 6000 flow sensor CW 6000 flow switch...

Страница 18: ...TY JS 20319 003 18 CW 6100 6200 flow sensor CW 6100 6200 flow switch...

Страница 19: ...TY JS 20319 003 19 CW 6260 flow sensor...

Отзывы: