27
A sudden change in the temperature of the examination room may cause moisture to build up in the wireless
detector. In the event of this happening, do not use the wireless detector until all of the moisture is removed.
Using the wireless detector with moisture in it may result in unstable system operation. If air-conditioning is
available, decrease the
temperature gradually to reduce the temperature difference between the wireless detector and the examination
room to prevent moisture build-up.
-
WEIGHT GUIDE
The wireless detector is not designed to support patients. Therefore it must be placed on a flat surface when
weight needs to be applied to its surface. Be careful not to damage the wireless detector by applying excessive
weight.
•
Even Distribution of Weight
Patients’ weight should be uniformly distributed across the detector’s surface. It should not be concentrated in
one place.
•
Localized Weight
When performing weight-bearing protocols, a patient should not stand on the detector without the ‘Weight
Distribution Cap’.
F
IGURE
8
Even Distribution of Weight
F
IGURE
9
Localized Weight (Center)
-
During Exposure
Do not apply excessive weight to the sensor unit. Otherwise, the sensor may be damaged.
The detector and patient must be taken from a fixed state. Otherwise, the performance may be
degraded.