background image

Preventative Maintenance Manual – Salute RDX 

2020/12/07

 

  1 

 

 

Preventive Maintenance 

Preventive maintenance should be carried out every 12 months 

Replace Air filter every 3 months. 
Replace Synchronous timer and valve set every 18 months. 

 

 

Grease the mechanical valve every 12 months. 

Check: 

a)

 

Case is free from debris 

b)

 

Compressor is seated properly in its housing 

c)

 

Electrical connections are secure and completely insulated 

d)

 

Pneumatic tubes are completely secure on compressor and rotary valve 

and are free from kinks 

e)

 

Wiring to printed circuit board (PCB) is intact 

f)

 

Outlet tubes and connectors are secure and fitted correctly 

 

Remove, service and re-install synchronous timer and rotary valve module 

a)

 

Identify the timer & valve module 

 

b) Disconnect pneumatic tubing from the three components - a. valve to compressor 
b. valve to solenoid/sensor c. valve to pump outlet 
                                                                                                         
                               

 

 

 

 

 
c) Unscrew and remove two screws.    Remove left and right metal fixtures. 

Отзывы: