16
Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
Figure below.
Fig 4-2: Equipment Connection for 2 signal capture
2) Set the frequency of one signal generator to 1800.4995MHz and the other to 1800.5005MHz. Set the output amplitude of both to
-20dBm. Respectively adjust the signal output amplitude of two signal generators and view the signals displayed on the spectrum
analyzer to make the displayed amplitudes of two signals equal.
3) Set up the spectrum analyzer to view signals.
Press
【
Preset
】
,
Press
【
Freq
】
,
Select [Center Freq] and input 1800[MHz],
Press
【
Span
】
key and select [Span], input 50[kHz];
Press
【
BW/SWP
】
, select [Res BW Auto Man], input 1[kHz].
At this moment, 2 signal peak can be seen on the screen of the spectrum analyzer but the 2 signals are not clearly separated, as shown
in Figure below
.
Fig 4-3: 1KHz RBW Trace