Figure 3-5 Test result of the FM modulation signal
3.3.3
Ф
M
For example: phase offset 180
°
, modulation rate 100kHz
Operation steps:
1. Instrument connection
Use a coaxial cable to connect the signal output end of the signal generator to the RF input port of the N9020A
spectrum analyzer.
2. Parameter setting
(1) Reset the signal generator
—
Press
【
Preset
】
button
(2) Parameter setting
—
Press
【
FREQ
】
button, input 1 GHz
—
Press
【
AMPT
】
button, input -10 dBm