List Sweep Output
S1131 supports list sweep function. User can edit the point freely and create a sweep list. This example creates a 4 point sweep
list.
1)
Connect S1131 to S3331 via a N type cable
2)
Preset the S1131, S3331
Press
【
PRESET
】
key . The instrument will return to default settings.
3)
Sweep settings
Press
【
Sweep
】
key.
4)
Select [SWP Mode], select [Freq+Ampl] in submenu.
5)
Select [SWP Type]. Select [List] in submenu, means list sweep
6)
Select [SWP Repe], select [Cont] in submenu, means continuous sweep
7)
Select [CFG List], click [Insert Row], add to 4 rows. Click [Goto Row], and enter 1, click [Confirm]. The selection box move
to the first row [freq]. Click [Edit Item], Set the [freq] to 500MHz. Then click [Goto Row] again and use the knob to move
the selection box to [Ampl], click [Edit Item] and set to -10dBm. Click [Goto Row], use the knob to move the selection box
to [Dwell] and set to 50ms.
Note: Knob can move the selection box when [Goto Row] is selected.
8)
Rotate the knob and the selection table will go to row 2. User can also use [Goto Row] to move the conversation box to