All enabled
Stimulation Set A disabled
Figure 8.22: (Left) Original current levels with all Stimulation Sets enabled. (Right)
Current levels with Stimulation Set A disabled.
8.13.3
Changing the ratio between Stimulation Sets
The ratio between Stimulation Sets can be changed, even during stimulation (Figure 8.23):
1.
Ungroup the Stimulation Sets (
).
2.
Select the Stimulation Set and change the current to the new desired level.
3.
Group the Stimulation Sets (
) and continue stimulation as normal.
Original ratios, Stimulation Set C selected
Stimulation Set C ratio lowered
Figure 8.23: (Left) A Program with 4 Stimulation Sets in Ungrouped mode with
Stimulation Set C selected. (Right) The ratio for Stimulation Set C has been reduced while
the others remain the same.
8.14
Neurophysiological Measurements
The CPA can be used with the eCLS/CLS to measure the neurophysiological characteristics
of the patients’ spinal cord as well as to test the vertical alignment of the two implanted
leads.
Caution: The Lead Offset program is currently in the testing phase of development
and should not yet be utilised.