Document Reference: ENG-UMAN-000731
Revision: 5.00
Page 24 of 61
5.
Navigate to the settings page by opening the Navigation Menu
and tapping
Settings (see Figure 8.6). Select the appropriate Patient Zone (see Figure 8.7).
6.
Return to the CLS programming page by opening the Navigation Menu
and
tapping Programming (see Figure 8.6).
Figure 8.6: Navigation Menu
Figure 8.7: Settings Menu with Patient Zone dropdown menu highlighted
8.5
Measure electrode impedance
Measure the electrode impedance to determine if all connections are intact between leads,
any lead adapters, and the stimulator.
1.
Touch the Impedance icon
.
Note: The icon will spin during the measurement and stop once it is complete.
2.
The impedance on the 24 electrodes and CLS case will be shown in the electrode
display window (see Figure 8.8).
Patient Zone
dropdown menu