Evoke
®
SCS System User Manual
15.1.6 Evoke Charger
Table 15-7: Evoke Charger specifications.
Materials
Charger
Control panel
Charger coil
ABS Plastic
PET
Silicone
Dimensions
Charger
139.7 mm x 82.6 mm x 26.2 mm
(5.50 in x 3.25 in x 1.03 in)
Charge coil
Cable length
60 cm (2 ft)
Diameter
10 cm (3.9 in)
Thickness
7 mm (0.28 in)
Weight
410 g (14.48 oz)
Communication
Inductive coupling
Link
Battery
2 x 18650 Lithium ion protected cells
Charging
Via the supplied power adapter
Battery Life
5+ years
Ingress
IP22 Rating for protection against access of solid objects greater than or
Protection
equal to 12.5 mm, and for vertically dripping water when the device is
tilted 15 degrees.
IEC 60601-1 /
Type BF Applied Part
EN 60601-1
Internally Powered Medical Electrical Equipment
Classification
Continuous Operation
Storage &
Temperature:
Min: -20 °C (-4 °F)
Max: 60 °C (140 °F)
Transportation
Humidity:
Min: 0% RH
Max: 90% RH
Conditions
Pressure:
70kPa (0.69 atm)
Max: 150kPa (1.48 atm)
Temperature:
Min: 5 °C (41 °F)
Max: 30 °C (86 °F)
Operating
Conditions
Humidity:
Min: 15% RH
Max: 90% RH
Pressure:
70 kPa (0.69 atm)
Max: 106 kPa (1.05 atm)
Supplied Power
Input Voltage
100-240 Vac, Class II Medical Electrical Equipment
Adapter Rating
Input Current
Max 0.3 A
Frequency
50 - 60 Hz
Output Voltage
8.4 Vdc
Output Current
Max 1.3 A
Ingress Protection
IP22 Rating
Note:
The Charger is isolated from the mains via the supplied power adapter.
CLIN-UMAN-005176
Revision: 4.01
Page 44 of 53