PC-150 Operation Manual
9
Setup Menu Mode /
Continued
SETUP MENU DESCRIPTIONS
NAME/CODE
DESCRIPTION
CODE/VALUE
F1
Display Mode
Selects display mode for the scale.
"0" = Non-NTEP mode
"1" = NTEP mode
0
√√√√
1
F6
Zero Range
Selects the range within which the scale may be zeroed. Note that
the scale must be in standstill to zero the scale.
100%
√√√√
1.9%
F9
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"0" = primary unit is lb. "1" = primary unit is in kg.
0
√√√√
1
F11
Display Check
Actuates the function that illuminates all digit segments, decimal
points, and LCD annunciators in a test sequence. Pressing the
ZERO
key to scroll down one level begins the test sequence.
Press
ZERO
key
to begin sequence
F12
Calibration
Places scale into the calibration routine. Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F13
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menus to
the default settings. USE WITH CAUTION!!!!!
Press the
ZERO
key twice to exe-
cute.
NOTE:
S
ub-menu F13 restores all parameters in the User (“A”) Menu and Setup (“F”) Menu
to the factory default settings. PLEASE USE CAREFULLY AS YOU MAY LOOSE VITAL
SETUP PARAMETERS!!!
Calibration
ON/OFF
SWITCH
DC POWER
JACK
RS-232
PORT
ID TAG
CALIBRATION
SWITCH COVER
CALIBRATION MODE KEY FUNCTIONS
ZERO
GROSS
NET
TARE
SET
lb
kg
The scale can be calibrated by following the procedure below. The minimum test weight that can
be used is 1.5 lb (0.7 kg), but 100 lb (50 kg) is the recommended test weight.