Salter Brecknell BS150 Скачать руководство пользователя страница 5

CAP.

CALIBRATION

WAY

Kg

SPAN

RANGE

CALIBRATION

WEIGHT

Lb.

SPAN

RANGE

CALIBRATION

WEIGHT

1/4

1000

5000

25 kg

910

4100

50 lb.

1/2

2000

10000

50 kg

1820

9100

100 lb.

100 kg

150 kg

Full

4000

20000

100 kg

4100

18000

200 lb.

1/4

1000

5000

100 kg

910

4550

200 lb.

1/2

2000

10000

250 kg

2270

11350

500 lb.

500 kg

600 kg

Full

5000

25000

500 kg

4550

22500

1000 lb.

1/4

1000

5000

250 kg

910

4100

500 lb.

1/2

2000

10000

500 kg

1820

9100

1000 lb.

1000

kg

1500

kg

Full

4000

20000

1000 kg

4100

18000

2000 lb.

1/4

1000

5000

500 kg

910

4550

1000 lb.

1/2

2000

10000

1000 kg

1820

9100

2000 lb.

2500

kg

3000

kg

Full

4000

20000

2000 kg

4550

22500

5000 lb.

4.  Press UNIT key to select weight unit: kg or lb

5.  When the weight is stable, press ON/ZERO/OFF key

to start calibrating.

6.  The display shows the calibration weight in normal mode.

7. Remove the weight from the platform and the calibration is

finished.

11

11.  Error message indication:

A.

Overload

B.:

Power on zero count too low

C.

Power on zero count too high

D. 

High zero count

E. 

Low zero count

F. 

Low battery indication

G. 

Eprom Error

12.  Two calibration weight units:  kg / lb

13.  Three calibration capacity: 1/4, 1/2, Full

14.  Power on zero-setting range: 

±

10%

15.  Zero range: 

±

2%

16.  Three software filter functions:  FAST/NORMAL/SLOW
17.  Three modes:  Normal mode/Setting mode/Internal Auto

Calibration mode

4

Err

0000

0000

Err

ErrL

--

ErrE

Содержание BS150

Страница 1: ...56031 Toll Free 800 637 0529 Voice 507 238 8702 Fax 507 238 8271 Email sales salterbrecknell com www salterbrecknell com February 2004 Copyright 2001 2002 Salter Brecknell Weighing Products Models BS1...

Страница 2: ...rt of the scale to locations where AC power may not be present Operation With only four keys the scale is simple to operate simply turn it on and begin weighing An automatic off function saves battery...

Страница 3: ...Features Portable operation or AC operation for full time use Each unit comes with an AC adapter for battery charging or full time use 14 hour charge time Stainless steel shroud Extra durable feet for...

Страница 4: ...LB and OZ unit COUNT Press count A 10 pcs Press count again for more choices of pieces to add When stable indicator is on place unknown amount on scale and count is functional Press UNIT key to return...

Страница 5: ...00 kg 3000 kg Full 4000 20000 2000 kg 4550 22500 5000 lb 4 Press UNIT key to select weight unit kg or lb 5 When the weight is stable press ON ZERO OFF key to start calibrating 6 The display shows the...

Страница 6: ...N PROCEDURE 1 Press ON ZERO OFF key without releasing and then press UNIT key 3 times to enter test mode the display shows zero count 1000 6000 2 Press ON ZERO OFF key and display shows 0 3 Load the c...

Страница 7: ...acity 3 6 15 30 60 150 300 600 1500 3000kg 4 OvL Overload 100 150kg 5 divS Division 1 1500 1 3000 1 6000 6 Filt Filter fast normal slow 7 CAL C Calibration Capacity 1 4 1 2 full 8 Hold Hold Average Ho...

Страница 8: ...6 Outline Drawing 7...

Страница 9: ...8...

Отзывы: