  OPERATION MANUAL 

   pH CONTROLLER

 

 

    

 

Please read the attached instructions carefully 

before use.

 

FRONT PANEL DESCRIPTION

 

 

1) Display 

  6) pH / SET function switch 

2) pH7 calibrate adjustment knob 

  7) HI / LO action function switch 

3) pH4 calibrate adjustment knob 

  8) Electrode BNC input terminal 

4) SET Adjustment knob 

  9) Control output socket 

5) Indicator of output in action 

10) AC power plug 

 

OPERATING (CALIBRATION)

 

 

1) Connecting AC power supply. 
2) Connecting the pH electrode into the pH input terminal socket. 
3) Turn the pH/SET switch to "pH" position. 
4) Put the pH electrode into the standard buffer solution pH7.00. Waiting a few 

seconds for stable reading and adjust the "pH7" knob until the display 
reading is exact on 7.00. Then take away the pH electrode from buffer 
solution, and clean the electrode with distilled water. 

5) Put the pH electrode into the standard buffer solution pH4.01, Waiting for a 

few seconds for stable reading and adjust the "pH4" knob until the display 
reading is exact on the buffer 4.01. (You also can calibrate pH10.00 by 
using the buffer solution pH10.00 instead of pH4.00 buffer solution.)   

(Make sure the step 4 to 5 are carried out when use a new pH 

electrode.

6) After calibrating the pH electrode, put the electrode into the water of tank. 

Turn the pH/SET function switch to “SET” position to adjust the setting point. 

7) Adjust the "SET" knob until the display reading exact meet your requirement. 

 
 

Slide the pH/SET function switch to "pH" position for measuring. (Slide the HI/LO 
switch to "HI" position will switch on the output control when pH measuring value 
goes above the setting value. Slide the HI/LO switch to "LO" position will switch 
on the output control when pH measuring value goes below the set value.) 

 

CLEANING AND MAINTENANCE 

 

1) The immersed level for electrode should not be higher than the electrode 

cap. (When use the refillable electrode, the level should be lower than 
filling open to avoid the water entering into the open and cause damage.) 
It had better cleaning the pH electrode sensor with soft brush and 
calibrating the pH electrode every 2-3 weeks to keep the test accuracy. 
Please keep the electrode in the storage bottle with 3M KCL buffer (or 
standard buffer solution pH4.0) solution, the crystallization is normal for 
the keeping.   

2) pH electrode is accurate glass product, please do not knock it, and do not 

use finger, cloth and alcohol to clean it to avoid the damage. 

3) There is protection "dead band" which will delay to switch on output control 

when the pH measuring value goes above 3-5 digits of the set value. The 
output control will switch off when the pH measuring value goes below 3-5 
digits of the set value. (HI/LO switch in "HI" position.)   

4) The output control will automatically turned off when slide the pH/SET 

function is switched in "SET" position to adjust the setting valve. 

5) Indoor uses only. 

 

GENERAL SPECIFICATION 

 

Power Supply 

:  AC100V - AC240V 50/60 Hz 

Power Consumption 

:  Approx. 2 watts. 

Display : 

0.56" 

LED. 

Measurement 

:  0 to 14 pH   

Resolution : 

0.01 

pH. 

Accuracy at 25°C 

:  ±(0.1%+ 2 digits) after calibrating. 

Impedance : 

10

12

 ohms.

 

Calibration Knob 

:  External pH7 (CAL.) & pH4/10 (SLOPE). 

Relay Contact 

:  5A for power supply voltage 

Control Range 

:  3.5 to 10.5 pH 

 

 

 

Control Output Voltage 

:  Same as plug in AC voltage. 

Temperature :  0 to 50°C (32 to 122°F). 
Operating Humidity 

:  Max. 90% RH. 

Dimension :  150 x 85 x 40 mm    (5.9 x 3.4 x 1.6 inch). 

 

WHY SHOULD WE KNOW pH RANGE 

 

pH value shows the acid or alkaline state of water (pH value equals 7 

is neutral, it is alkaline as pH value is greater than 7, and acid as pH 

value is less than 7), most of the fish can adapt to life in water around 

pH =7, but many species have special water preferences.   

Such as: 

species of pH < 7: Discus, Amazon cardinals, preferences for 

pH=6.0 to 6.5. 

species of pH

7: Marine fish preferences for pH

8.4 and 

African cichlids preferences for pH 7.5 to 8.0

 

 

THE RIGHT pH RANGE KEEP THE FISH RIGHT

 

 

WHAT TO DO AS pH RANGE IS TOO LOW 

 

The pH range of water should be shifted up, the methods of 

adjustment you can take: Use coral gravels substrate, anion 

exchange resin, add additives of powder, solution or tablets which 
can shift pH

KH and GH range up.   

 

WHAT TO DO AS pH RANGE IS TOO HIGH 

 

The pH range in water should be shifted down, the methods of 

adjustment you can take: Use weak-acid gravels substrate, cation 

exchange resin and higher portion of reverse osmosis water. Add 
additives of powder or solution which can shift pH range down. 

 

Made in Taiwan

 