PRELIMINARY OPERATIONS
1 - Connect instrument to GND
2 - Finger on touchplate
3 - Hold Sensor with hand
4 - in the MONITOR MENU
Verify ZERO on display
(with WARM-UP ALARM still active you could have values starting from
-30V, offset values between -0005V and +0005V are to be considered in tolerance )
MONITOR MODE
In this menu if the operator press
<START/STOP>
the
instrument shows the electric potential present on the
probe.
The electric potential is mediated every second.
A beep every second indicates that the instrument is
measuring. The beep was made to avoid missing the
instrument with measuring active, causing battery
consumption.
To deactivate the measuring press
<START/STOP>
again or
<EXIT>
.
MONITOR
(Datalogger OFF)
DATALOGGER MODE
If there is free memory space the instrument can log the
graph of a new walking test.
In this menu if the operator press
<START/STOP>
the
instrument shows a pop-up window with a short countdown
(3 seconds) before to execute the test.
In the pop-up window is indicated the maximum duration of
the walking test.
To stop the data acquisition of the graph press
<START/STOP>
.
DATALOGGER
Graph n.3
+Peak=26V
-Peak=-6
Press
<LEFT>
or
<RIGHT>
to scroll the graph for a quick evaluation.
Press
<START/STOP>
again: a new pop-up window asks you to save or discard the graph.