- 35 -
7.
Maintenance and repair
Repairs may only be carried out at the manufacturer's factory!
Inspection intervals according to DGUV regulation 3 must be observed!
Exchange of the internal fuses for the electronics module and the shockproof socket:
If the controller module has no function, the fuse for the controller unit may be
defective.
If the temperature regulator is working, but the connected consumer has no
function, the fuses of the shockproof socket can be defective.
The fuse holder for the controller unit is located on the rear of the controller. The
caps for the power fuses are located in the safety socket, which can be opened with
a screwdriver.
Replace the fuse with new fuses of the same value (controller module 1x 20 x 5 mm,
1 A fast acting; shock-proof socket 2x 20 x 5 mm, 10 A slow-blow).
Installation in reverse order.
Note on the stand clamp:
A stand clamp is located on the back of the controller to clamp support rods up to Ø
12 mm or to attach the controller to a stand.
This support clamp cannot be removed, as it also serves as a support for the
internal structure of the controller.
Содержание LabHEAT KM-RX1001