E l e c t r o n i c   l a b o r a t o r y   r e g u l a t o r  

t y p e  

KM-RX1001 

 

 

 

 

 

 

 

O P E R A T I N G     M A N U A L

 

 

 

 

 

Содержание LabHEAT KM-RX1001

Страница 1: ...Electronic laboratory regulator type KM RX1001 O P E R A T I N G M A N U A L...

Страница 2: ...onnection as well as operation parameterization and configuration of the device For further information please refer to the operating manual B 702030 0 available as PDF document This manual also conta...

Страница 3: ...g personnel 12 2 4 Safety instructions for maintenance 13 3 Product description 14 3 1 Execution and technical data 14 3 2 USB interface 15 4 Assembly and connection 16 4 1 Connection diagram 16 5 Com...

Страница 4: ...ation ConF 23 6 2 1 Analog input 24 6 2 2 Controller 25 6 2 3 Ramp function firing curve 26 6 2 4 Limit value monitoring 27 6 2 5 Timer 29 6 2 6 Outputs 30 6 2 7 Binary function 32 6 2 8 Display and o...

Страница 5: ...n on copyright and data protection This operating manual must be treated confidentially It should only be made available to authorized persons It may only be made available to third parties with the w...

Страница 6: ...warnings must be strictly observed Observe the safety instructions mentioned in each case and be particularly careful in these cases Pass on all safety instructions to other users In addition to the i...

Страница 7: ...Scope of delivery Check that the scope of delivery is complete and undamaged Contact your supplier immediately if parts are missing or defective 1 x Laboratory regulator type KM RX1001 1 x Operating...

Страница 8: ...al damage The product must not be installed in the Ex area If process variables from the Ex area are processed with the product installed outside the Ex area all supply lines of the product that lead...

Страница 9: ...al in a safe place for later use The regulations mentioned in this operating manual apply in all EU countries When using the device in a country outside the EU the relevant national regulations must b...

Страница 10: ...operating manual must be observed The operating instructions of the operator must be observed The statutory accident prevention regulations must be observed The following is considered improper opera...

Страница 11: ...t prevent these residual risks from leading to accidents or damage Danger Penetration of moisture by water or chemical liquids can cause dangerous to life threatening electric shocks If moisture gets...

Страница 12: ...g manuals All faults and especially those that could impair safety must be eliminated immediately The product must not be operated unattended Any person entrusted with the installation commissioning o...

Страница 13: ...egular intervals for safety and proper functioning in accordance with the applicable regulations Only testing in accordance with DGUV V3 is permitted Danger Repairs to the product can cause serious in...

Страница 14: ...ated on the rear side as a clamping device for attaching the temperature regulator to a stand 3 1 Execution and technical data Nominal voltage 230 V 50 Hz Nominal power max 2300 Watt resistive load Co...

Страница 15: ...u have the possibility to connect the control insert to your computer via a USB A to Mini B plug cable The software for configuration and parameter setting via computer can be downloaded via the follo...

Страница 16: ...nd other inductive loads The ambient temperature at the installation location may be 0 50 C at a relative humidity of 75 without condensation Aggressive gases and vapors can destroy the device The ins...

Страница 17: ...adapted to the respective consumer and must be connected in series before the temperature regulator The maximum fuse rating is 16 A A cable cross section of at least 1 5 mm is recommended Set the mai...

Страница 18: ...or 3 RCD adapter 5 2 Adaption The product is delivered pre configured with the basic settings described in the operating manual The operator must adapt the product to the controlled system of the plan...

Страница 19: ...ge other parameters Due to the PID behavior of the controller the ACTUAL temperature usually oscillates well above the adjusted SET temperature during the first heat up By activating self optimization...

Страница 20: ...tion configuration and parameter setting 6 1 Operation 6 1 1 Display and operating elements 6 1 2 Self tuning setpoint and manual operation Starting from the normal display the following functions are...

Страница 21: ...21 6 1 3 Ramp function firing curve 6 1 4 Level concept The parameters for setting the controller are organized in different levels The different levels are also accessible in manual mode...

Страница 22: ...22 Level locking all levels are free at the factory 6 1 5 Operator level OPr Depending on the configuration the following parameters are available...

Страница 23: ...den if the required equipment is not available Factory settings are shown in bold in the following tables Analogue sector For some parameters at the configuration level the user can select from a rang...

Страница 24: ...24 6 2 1 Analog input...

Страница 25: ...25 6 2 2 Controller The controller receives the actual value from the analogue input...

Страница 26: ...26 6 2 3 Ramp function firing curve The device can be operated as a fixed value controller with or without the ramp function A firing curve is also supported...

Страница 27: ...7 6 2 4 Limit value monitoring The device has two functions for limit value monitoring Li1 Li2 each with 8 different alarm functions AF1 to AF8 Both output signals are available for the binary functio...

Страница 28: ...28...

Страница 29: ...ignal which is available for the binary functions Time limited control or a time dependent target value switching can therefore be realised The timer value is not ensured during a power failure The ti...

Страница 30: ...OutA The switching states of the binary outputs 1 to 3 5 are represented by the LEDs K1 to K3 K5 LED lights up output active Binary outputs The instrument has one relay output make contact output 1 as...

Страница 31: ...31 Analogue output The device can be optionally equipped with an analogue output output 2...

Страница 32: ...be triggered with the binary signals from the binary inputs limit monitoring and timer The binary functions for starting and stopping react to the rising edge of the triggering signal All other binar...

Страница 33: ...33 6 2 8 Display and operation Both displays and the function key can be individually adapted to the respective requirements...

Страница 34: ...34 6 3 Parameterization PArA...

Страница 35: ...socket can be defective The fuse holder for the controller unit is located on the rear of the controller The caps for the power fuses are located in the safety socket which can be opened with a screw...