S.E. International URSA II Скачать руководство пользователя страница 30

30

Generally speaking, most adjustments to expand the separation between peaks should be performed using 

the HV adjust. Start increasing the gain only if the peak separation remains too small even at the maximum 

HV recommended for the particular detector; increase the gain until the peak separation is slightly wider than 

wanted, then reduce the HV as needed. If the peaks are in a good location but appear distorted, try lowering 

the HV and increasing the gain.
Adjust the threshold lower if too much of the low energy portion of the spectrum is being cut off. Adjust the 

threshold higher if noise is apparent in the lowest regions of the spectrum.
In using the hardware setting adjustments, the associated “Set” button will become enabled, indicating that 

the value has been changed from those at which URSA-II is currently set. For example, if you change the HV 

the “Set HV” becomes enabled. Clicking the associated “Set” button will (1) pause the current acquisition if 

one is occurring, (2) set all hardware settings that have been changed, not just the one the particular button 

is associated with, and (3) restart the acquisition using the new settings if one had been occurring without 

clearing the previous spectrum.
Although it is completely up to the user (and subject to the characteristics of the detector being used), it has 

been generally found that adjusting the hardware settings such that the 662 keV peak is at about channel 1200 

will give an overall energy range coverage to about 2,300 keV.
Hardware settings are automatically saved and applied to a specific URSA-II/detector combination any time a 

“Set” button is pressed. If different sets of hardware settings are desired for the same detector, create a “new” 

detector based on that detector, changing the description to reflect the different purpose.

Procedure

4. Click on the “Hardware” button on the URSA-II button panel. This will open the hardware settings panel 

along thebottom of the spectrum display.
5. Check the “Allow Edit” check box to enable the hardware settings.
6. Ensure that the HV is set to a reasonable value. For relatively new URSA-II users, coarse gain should be at x2 

(thelowest setting) and the fine gain should be at x0.500 for an overall gain of x1; the threshold should be at 50 

mV.
7. Click the “Start Acquisition” button to begin counting the source. Allow the source to count long enough to 

see therelative locations of the two 137Cs peaks. NOTE: If excessive noise is clearly visible in the lowest portion 

of thespectrum, increase the threshold, clear the spectrum, then restart acquisition. Repeat this process as 

needed.
8. If no peaks are visible, increase the HV and go back to step 4.
9. If both 137Cs peaks are not visible:
10. If there are very few counts to the right of the visible peak but many to the left (i.e., only the 662 keV peak 

is visible) and it is towards the right of the spectrum display, increase the HV and go back to step 4.
11. If there are many counts to the right of the visible peak (i.e., only the 32 keV peak is visible and the 662 keV 

peak is out of view to the right), decrease the HV and go back to step 4.
12. If there are very few counts to the right of the visible peak but many to the left (i.e., only the 662 keV peak 

is visible) but the peak is in close to the desired location on the spectrum, adjust the threshold closer to zero 

and go back to step 4.
13. If both 137Cs peaks are visible, but are farther apart than wanted, decrease the HV and go back to step 4.

Содержание URSA II

Страница 1: ...1 S E International Inc P O Box 39 436 Farm Rd Summertown TN 38483 USA 1 800 293 5759 1 931 964 3561 Fax 1 931 964 3564 www seintl com radiationinfo seintl com INTERNATIONAL OPERATION MANUAL...

Страница 2: ...s 9 Misc 9 View Menu 9 Sound 10 Library 10 Transfer 10 Help 10 Button Bar 11 Hardware Status Bar 12 New User Hints Bar 12 Spectrum Display 12 Spectrum Status Bar 12 Acquisition Controls and Indicators...

Страница 3: ...erforming an Energy Calibration 26 Performing a Shape Calibration 28 Creating a Regions of Interest ROI Set 29 Manually Creating or Editing an ROI Set Manual Method 30 Performing an Efficiency Calibra...

Страница 4: ...URSA II accepts positive negative or pre shaped positive pulses Pulse heights should not exceed approximately 5 volts Uses 110V or 220VAC to 12V DC adapter 1 25A or runs on 6 internally recharged NiM...

Страница 5: ...ined batteries voltages The NiMH batteries should power the URSA II for about four hours Note however that this figure can change dramatically based on the current draw of the detector high voltage ot...

Страница 6: ...in and power on URSA II 4 Start the software URSA II MCA searches all COM ports for the URSA II automatically 5 A detector must be entered before URSA II can be opened Many of URSA II s calibration f...

Страница 7: ...ing a displayed spectrum to an URSA II spectrum file usf The next two Save Spectrum and Load as Background and Save Spectrum and Load to Static Display also save the spectrum but additionally either l...

Страница 8: ...order to save printer ink This is the recommended method of printing charts The final menu item Exit closes URSA II MCA Warnings will occur if the currently displayed spectrum has not been saved and...

Страница 9: ...n See instructions on page 32 The next seven QQQ menu items Full Screen through Sample Data Panel are identical in function to the seven buttons in the Button Panel on the lower right of the Spectrum...

Страница 10: ...h nn ss where yyyy is the four digit year mm is the two digit month dd is the two digit day of the month hh is the two digit hour in 24 hour format nn is the two digit minutes past the hour and ss is...

Страница 11: ...se common sense in applying different colors For example if the Background color is set to be the same as the Main Spectrum color the display will appear blank Undo Zoom gives a method for restoring t...

Страница 12: ...button caption becomes Hide and clicking on it or on the x in the upper right corner will close the window The order of the display can be based on energy or yield ascending or descending Change this...

Страница 13: ...area Once a spectrum is zoomed it can be panned by rightclicking the mouse on the spectrum display and dragging it Un zoom by dragging a box any size anywhere on the spectrum display from lower right...

Страница 14: ...isition has previously been stopped The Clear button erases all data from the spectrum and resets the elapsed time to zero If a spectrum is currently acquiring acquisition will stop No warning is give...

Страница 15: ...when the spectrum is displayed in 256 or so channels URSA II MCA s 4096 channels of data can be filtered to give a smoother representation of the energy continuum Try using different values of smoothi...

Страница 16: ...nudges to compensate for spectrum drift caused by temperature changes in the detector If an even smaller shift in gain is needed try adjusting the high voltage by tiny increments including decimal val...

Страница 17: ...set This name becomes the one displayed in the spectrum status bar and is also used to identify individual ROIs in analysis reports Approximately in the middle of the ROI panel are two radio button co...

Страница 18: ...will be unlocked for editing If this box is not checked the data for the last calibration can be viewed but not altered The arrow buttons will change the currently selected specific ROI within the gre...

Страница 19: ...ed printed or saved by clicking the Report button Note The contents of the report and use of the report window are discussed in detail in other portions of this manual Peak Based This displays the Pea...

Страница 20: ...h and this method of adding found peaks may beused instead of or as a supplement to the computerized search When Show FWHM is checked the FWHM full width at half maximum calculated for each found peak...

Страница 21: ...sults will be displayed in subsequently displayed reports The Save Spectrum button repeats the function under the File Menu It is also included here for convenience The Clear Fields button deletes the...

Страница 22: ...emoved to improve performance Both libraries contain many calculated multiplets as an aid These can of course be edited or disabled Either library can be edited and saved under a different name creati...

Страница 23: ...ently displayed isotope Click on any photon energy in the grid and use the Kill Energy button to completely delete that photon energy from the currently displayed isotope Use the Add Isotope button to...

Страница 24: ...g or recording changes in count rate over time An analog meter style display also shows the current count rate MCS mode allows you complete control over how the data is accumulated and displayed Menu...

Страница 25: ...rt trace to be displayed as a line as points as a line with points the default bars or as an area graph The next three menu items Chart Background Color Chart Trace Color and Chart Labels Color allow...

Страница 26: ...unt of time in the input box Typing in the Time box will immediately un check it pressing Enter on the keyboard after changing the Time value will re check it Directly underneath the Time sec input bo...

Страница 27: ...erform Energy Calibration 3 Perform a Shape Calibration Following this step URSA II is ready to provide qualitative analysis 4 Create a Regions of Interest ROI Set This is in preparation for efficienc...

Страница 28: ...e URSA II has two detector inputs Input 1 is the series C connector right next to the DB9 serial connector This input has the signal and the high voltage coupled just like on most hand held instrument...

Страница 29: ...it can be displayed usefully as a spectrum These two controls are used together to create an overall gain The overall gain is generally the value you should follow It is worth mentioning that the gai...

Страница 30: ...ed If different sets of hardware settings are desired for the same detector create a new detector based on that detector changing the description to reflect the different purpose Procedure 4 Click on...

Страница 31: ...that detector is attached to the URSA II the URSA II is switched on the URSA II MCA software has been successfully started a radioactive source of known isotopic composition is on the detector and URS...

Страница 32: ...simple spectrum such as 137Cs 2 From the Utility menu select Energy Calibrate The energy calibration window will appear If this is the first energy calibration for this detector the energies for the...

Страница 33: ...een performed Discussion For NaI Tl detectors the Full Width at Half Maximum FWHM is proportional to the square root of the centroid energy as is the overall peak area The Shape Calibration helps dete...

Страница 34: ...t method a spectrum is acquired and the Find Peaks buttons is pressed from the Peak Based panel edited as desired then saved as an ROI set The ROI set can be further edited from the ROI Panel includin...

Страница 35: ...while the single arrows adjust the settings in approximately 0 5 keV increments c To delete a peak first select the peak to be deleted using the Selected Peak arrow set Once the peak is highlighted cl...

Страница 36: ...io button c Position the cursor at the suitable location on the spectrum display and double click d The energy and associated channel will be displayed in the appropriate boxes on the ROI Panel e When...

Страница 37: ...sed Procedure 1 Acquire a spectrum of your efficiency calibration source A background spectrum should also be subtracted from the display if that is how your samples will be analyzed 2 Click on Effici...

Страница 38: ...small window will pop up and display an isotope selection drop down menu and an energy peak selection drop down menu The hardware settings panel will also open along the bottom of the spectrum display...

Страница 39: ...ts suppliers be liable for any consequential incidental direct indirect special punitive or other damages whatsoever including without limitation damages for loss of business profits business interrup...

Страница 40: ...or will repair the defect and return the instrument to you prepaid without charge for parts or labor Shipping cost to the factory for repair is the responsibility of the customer Customer is also resp...

Страница 41: ...S CITY STATE ZIP COUNTRY Please fill out this form and send it back to us if you would like to be notified of the NIST calibration renewal for your instrument to S E International Inc P O Box 39 436 F...

Отзывы: