Error description
Possible cause
Explanations and action taken
Temperature display still
shows 0 / 100 °C
Contact between cell and
detector block broken
Make sure connector is plugged in
Temperature sensor within the
measurement cell defective
Replace measurement cell
Unstable or noisy
basis line
Bubbles in the solvent
Use degassed solvent
Measurement cell soiled
Rinse out the measurement cell with a
cleaning solution (see above) and switch
the equipment off and then on again.