39
The specifications and instructions from the manufacturer re-
ferring to cleaning and disinfectant materials and those
referring to the corresponding devices must be followed.
6.3.1
CLEANING THE MOTOR UNIT
Do not use disinfecting sprays. Dampness could penetrate
and lead to bearing damage.
We recommend RUCK alcohol-free disinfectant cloths (item no
2942801) for gentle cleaning and disinfection.
The holder can be disassembled for cleaning.
To clean the inside of the grip sleeve, it can be pulled off by gently tilt-
ing the motor casing and motor base apart.
This releases the connection with the motor.
The motor and chuck can then be slid down and out, for example using
the drill pin supplied.
The inside of the grip sleeve can then be cleaned.
To reassemble the holder, align the motor unit with the guide rails on
the motor casing. When assembling the motor connection, please at-
tach and slide in the plug connectors (02) carefully to avoid damaging
the contact pins (01). Make sure the plug connectors are correctly and
symmetrically aligned.