PortaMonitor
Page 8-2 Calibration and Service
8.3. Open the unit
Important:
This description is not intended to be used as a service
manual. These are guidelines for an expierenced service engineer to
proceed adjustments or exchange parts of our instrument. However,
we strongly recommend repair of faulty units in our factory.
Please refer to the safety instruction in chapter 6 before continuing.
1. Disconnect all cables from the PortaMonitor.
2. Place the PortaMonitor face down on the work bench. Asure not to
scratch or damage the display.
3. Open the backplane cover (four screws). Store screws and metal
plate. (fig. 8-2)
8.4. Display exchange
In case of a faulty display this has to be exchanged completly including
the mainframe. Special tools would be required to mount a display in
the frame. Thats why only preproduced display units are available as a
spare part. Proceed as follows to exchange display:
1. Disconnect all cables from the PortaMonitor.
2. Place the PortaMonitor face down on the work bench.
3. Open the backplane cover (four screws). Store screws and metal
plate. (fig. 8-2)
4 S c h r a u b e n
4 s c r e w s
1 .
A
B
2 .
Fig. 8-2: Remove rear cover
4. Remove an optional table stand.
All manuals and user guides at all-guides.com
Содержание PortaMonitor 1061
Страница 2: ...All manuals and user guides at all guides com...
Страница 34: ...PortaMonitor Seite 4 8 Kalibrierung und Service All manuals and user guides at all guides com...
Страница 60: ...PortaMonitor Page 8 8 Calibration and Service All manuals and user guides at all guides com...
Страница 70: ...PortaMonitor Seite Page 9 10 Zeichnungen Drawings All manuals and user guides at all guides com...
Страница 74: ...PortaMonitor Seite A 4 Technische Daten All manuals and user guides at all guides com...
Страница 78: ...PortaMonitor Page B 4 Specifications All manuals and user guides at all guides com...