background image

Chamber Adapter User's Manual

2019-10-14/1.6A

Function and connection

12

2

Function and connection

2.1

Function of the Chamber Adapter

The Chamber Adapter is an ion chamber adapter with a built-in bias voltage supply for use with ion chambers
that have a Lemo connector and operate on negative bias voltage, +300 V DC.

The Chamber Adapter is powered ON with the black button to the right. 

The green power LED indicates that it is on.

The red LoBat LED indicates when it is low battery and charging is required. 
Note:  Do not measure when the LoBat LED is on since it may lead to inaccurate
measurements.

The yellow Bias LED indicates that +300 V is present.  The bias is activated
when the Chamber Adapter is turned on.  The bias is turned off when the
Chamber Adapter is turned off.

The Chamber Adapter has a rechargeable battery.  The Piranha power supply (not the Cobia power supply)
can be used to charge the Chamber Adapter.  To charge the Chamber Adapter:

1.  Turn off and disconnect the Chamber Adapter from the Piranha.

2.  Connect the Piranha power supply to the Chamber Adapter and make sure that the main switch is in the
off position and the Charge LED is on.

3.  When the battery is fully charged the Charge LED goes off.

2.2

Connecting the Chamber Adapter

Make sure the Chamber Adapter is powered off before connecting the ion chamber.

1.  Connect the ion chamber to the input on the Chamber Adapter (Lemo connector). 

2.  Power on the Chamber Adapter.  The bias is now applied to the ion chamber.

3.  Connect the Chamber Adapter to the Piranha or Cobia.

4.  Make the necessary settings/selections in Ocean 2014 or in QABrowser.  The sections "Measurement with
Ocean 2014" and "Measurement with QABrowser" contain more detailed instructions on how to use the
Chamber Adapter in Ocean 2014 and in QABrowser.

Содержание Chamber Adapter

Страница 1: ...Chamber Adapter User s Manual English Version 1 6A RTI article number 4700 000784 Chamber Adapter Ion chamber adapter for Piranha...

Страница 2: ...Chamber Adapter User s Manual 2019 10 14 1 6A...

Страница 3: ...Chamber Adapter The Chamber Adapter makes it possible to use ion chambers with the Piranha and Cobia...

Страница 4: ...f the product without prior permission from RTI Group AB RTI Electronics Inc 33 Jacksonville Road Building 1 Towaco NJ 07082 USA Phone 800 222 7537 Toll free Int 1 973 439 0242 Fax Int 1 973 439 0248...

Страница 5: ...connect an ion chamber that is not intended for 300 V bias voltage to the Chamber Adapter The Chamber Adapter is only intended to be used with ion chambers that has a LEMO connector or with available...

Страница 6: ...Chamber Adapter User s Manual 2019 10 14 1 6A...

Страница 7: ...Function and connection 2 12 Function of the Chamber Adapter 2 1 12 Connecting the Chamber Adapter 2 2 13 Non RTI ion chambers 2 3 13 Adding a new ion chamber 2 4 17 Application example 3 18 Measurin...

Страница 8: ...Chamber Adapter User s Manual 2019 10 14 1 6A...

Страница 9: ...Chapter 1 Introduction...

Страница 10: ...d for other applications For the purposes of this documentation the expression electrometer refers to the electrometer of the Piranha or Cobia The Chamber Adapter makes it possible to choose between s...

Страница 11: ...Chapter 2 Function and connection...

Страница 12: ...apter has a rechargeable battery The Piranha power supply not the Cobia power supply can be used to charge the Chamber Adapter To charge the Chamber Adapter 1 Turn off and disconnect the Chamber Adapt...

Страница 13: ...TNT Adapter Cable 9730099 01 Malco to Lemo Adapter Cable You cannot connect ion chambers with a BNC Banana connector 2 4 Adding a new ion chamber The calibration factor for the preferred ion chamber...

Страница 14: ...BCA is the chamber adapter identification information This information is locked at the factory during manufacturing and cannot be changed If you click on a detector to the left the available calibrat...

Страница 15: ...x 9 When you click OK to continue a new pop up window is shown where you can select the calibration details Select the calibration you want from the drop down list for our example we will choose Mo Mo...

Страница 16: ...nformation into the device Do not interrupt the program before this process is completed 13 Quit the Detector Manager when the new data is stored in the device As soon as you start Ocean 2014 or the Q...

Страница 17: ...Chapter 3 Application example...

Страница 18: ...pear The detector types that are available with the Chamber Adapter are identified by the text PBCA Piranha Bias Chamber Adapter If you are running Ocean 2014 in Keyboard mode all of them are availabl...

Страница 19: ...tor Manager see Adding a new ion chamber section your own name is shown here instead 6 Select the ion chamber you want to use and click OK 7 The measurement is activated and you can start to make expo...

Страница 20: ...on the Chamber Adapter 4 Activate the measurement CT Dose for example 5 The detector selection will appear Your particular detector selection picture may be different than those shown in this manual...

Страница 21: ...nal button There are two drop down menus shown in the next window The upper menu shows the available probe types and when selected the lower part shows the serial number of the probe that is selected...

Страница 22: ...Chamber Adapter User s Manual 2019 10 14 1 6A Application example 22...

Страница 23: ...Chapter 4 Specifications...

Страница 24: ...e Ripple 15 mV Power source Rechargeable battery Battery life 10 hours Current range 10 pA to 0 1 A 2 or 1 pA exp time 10 ms Air kerma rate DCT10 0 8 mGycm s 8 Gycm s 5 or 0 08 mGycm s exp time 10 ms...

Страница 25: ...Chapter 5 FAQ...

Страница 26: ...ang if it gets a very high signal for example when connecting or disconnecting the chamber I get the wrong time when measuring short times The Chamber Adapter is not suitable for time measurements esp...

Страница 27: ...Applications CT 10 Mammography 10 B Battery 12 24 Bias 12 BNT 13 C calibration factor 13 Chargeing 12 D Detector Manager 13 I Internal detector 20 Ion chamber 13 calibration factor 13 connect 12 L Le...

Отзывы: