21
CMRR / NMRR : (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
AC
: CMRR > 60dB at DC, 50Hz / 60Hz
V
DC
: CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type :
71 :
Average sensing rms indication.
72 / 73 :
AC conversions are ac-coupled true rms responding, calibrated to the
sine wave input.
*The basic accuracy is specified for a sine wave below 4000 counts. Over 4000
counts, add 0.6% to the accuracy. For a non-sine wave, the crest factor of the
waveform is specified at
≦
3 at full scale up to 2000 counts, decreasing linearly to
a crest factor
≦
1.5 at 1000 counts. Add
±
1.5% for a non-sinusoidal waveform.
Crest Factor (C.F.) is the ratio of Peak value to RMS value.
(2) DC / AC Current
Overload Protection :
A input : 10A ,600V, high energy fuse. (Not IDM 71)
µA input : 600V rms.
* 1 AC Conversion Type :
Conversion type and additional specification are same
as DC/AC Voltage.
*2 (Not IDM 71)
Range
DC Accuracy
AC Accuracy
Voltage
Burden
600.0µA
±(
1.0% + 2 dgt)
N/A
<4mV / µA
6000µA
6.000A *2
2V max
±(
1.5% + 5 dgt)
50Hz ~ 500Hz
*1
10.00A *2
Specification