RS PRO 877-1791 Скачать руководство пользователя страница 3

RS, Professionally Approved Products, gives you professional quality parts across all products categories. Our range has 
been testified by engineers as giving comparable quality to that of the leading brands without paying a premium price. 

Products Details 

 

 

 

 

  

Measuring Force: 

 

5 – 10N 

Power Consumption: 

 

Greater than 35 milliamps 

Operating Temperature: 

 

0 – 40 deg.C 

Storage Temperature: 

 

-20 to 60 deg. C 

Protection Class: 

 

IP54 (resistant to water splash)   

  

Operating Care

 

Clean measuring faces with a clean soft cloth only 
Do not use any organic solvent for cleaning such as acetone etc. 
Keep instrument away from strong magnetic fields and high voltage environments which can affect the correct 
working of the electronic pack 
Prevent the ingress of oil and liquids into the electronics 
Do not use or store the micrometer in direct sunlight, or in an excessively hot or cold environment 
Remove battery if the instrument is not to be used for a long period of time 
Do not disassemble or drop the instrument 
  

Do not mark the instrument by engraving, etching or any other permanent method of marking as this will 
invalidate the warranty 

Fault Finding 

Failure 

Causes 

Remedy 

Display: “E       1” 
Display: “Exxxxx” 

Measured value is over display range 

Reset the origin or change to relative 
mode 

Display: “E       2” 

The origin is too great 

Reset the origin 

Display: “E       3” 
Display: “E       8” 

1 The micrometer is disturbed 
2 Something wrong with sensor 

1 Reset the battery 
2  return the micrometer for repair 

Measured value is not correct 

1 Measuring surfaces are not clean 
2 The origin is incorrect 

1 Clean measuring surfaces 
2 Reset the origin 

Display is confused or dead 

Strong disturbance to micrometer 

Reset battery 

No display 
Display is blurring 
Battery sign appears 

Battery voltage below 1.45V 

Replace battery 

RS, Professionally Approved Products, gives you professional quality parts across all products categories. Our range has
been testified by engineers as giving comparable quality to that of the leading brands without paying a premium price

.

ENGLISH

Содержание 877-1791

Страница 1: ...etting Rods Supplied in fitted case Packed Weight and Dimensions RS Code Description Weight g W mm H mm L mm 8771791 Electronic Micrometer 0 150mm 0 6 1960 265 45 425 RS Code Range mm 1nch A mm L mm Accuracy mm 8771791 0 150 82 158 0 003 Power 1 x SR44 1 5V battery 2 Locking Device 3 Thimble 4 Ratchet Stop 5 ON OFF SET Button 6 ABS INC UNIT Button 7 Data Output button not available 8 Battery cover...

Страница 2: ...t until required number is displayed 7 Press and hold SET button to advance to the next digit 8 Repeat actions 6 7 until all the digits on the display have been programmed 9 Once the last digit has been programmed press and hold the SET button until Set flashes in the display 10 Press and release SET button to cancel the flashing Set on the display 11 The micrometer is now set and ready for use Bu...

Страница 3: ...s not to be used for a long period of time Do not disassemble or drop the instrument Do not mark the instrument by engraving etching or any other permanent method of marking as this will invalidate the warranty Fault Finding Failure Causes Remedy Display E 1 Display Exxxxx Measured value is over display range Reset the origin or change to relative mode Display E 2 The origin is too great Reset the...

Страница 4: ...ge block to check the zero position If the zero point does not line up as required it can be corrected by using the following procedure When the zero point deviation on the thimble is under 2 divisions from the graduated base line Turn the sleeve using the C spanner provided until correct alignment is achieved When the zero point deviation on the thimble is over 2 divisions from the graduated base...

Отзывы: