RS PRO 205-0963 Скачать руководство пользователя страница 6

page 6 of 24 

Operating Manual RS 1107, RS 1113 

 

rspro.com

 

2.3 

Safety instructions 

 

 

NOTE 

 

This product does not belong in children's hands! 

2.4 

Intended use 



The device is designed as a manometer and measures even the smallest pressure 
differential pressures of up to ±200 hPa with a maximum resolution of up to 0.1 Pa in 
air or in non-corrosive/non-ionising gases between the two pressure connections. 

 

 

RS 1107

 

RS 1113

 

Differential pressure 

±200 hPa 

±2000 hPa 

Max. resolution 

0.1 Pa 

1 Pa 

Usual applications include precise measurements of filter condition, gas flow pressure, 
draught, leak integrity, dynamic pressure flow speed. 

The pressure connection is made at the supplied interchangeable pressure connection 
ports with suitable hoses. 

The product must only be used under the conditions and for the purposes for which it 
was designed. 

It must be handled with care and used according to the technical data (do not throw, 
strike, etc.). Suitable measures must be used to protect the pressure connections and 
be protected from dirt and moisture. 

 

 

Содержание 205-0963

Страница 1: ...H88 0 2X DK2RS 1 0 Operating Manual EN RS 1107 UT Stock number 205 0963 RS 1113 UT Stock number 205 0964 rspro com ...

Страница 2: ...glance 7 3 1 The RS 1100 manometer series 7 3 2 Display elements 7 3 3 Operating elements 8 3 4 Connections 9 4 Operation 10 4 1 Opening the configuration menu 10 4 2 Open the adjustment menu 12 5 Measurement Basics 13 5 1 Special functions 13 5 1 1 NVLL Tare function 14 5 1 2 F NE High resolution measurement with 0 1 Pa RS 1107 14 5 1 3 F NE High resolution measurement with 1 Pa RS 1113 15 5 1 4 ...

Страница 3: ...ting Manual RS 1107 RS 1113 page 3 of 24 6 2 1 Battery indicator 17 6 2 2 Changing battery 17 7 Error and system messages 18 8 Technical data 20 8 1 RS 1107 20 8 2 RS 1113 22 9 Service 24 9 1 Manufacturer 24 ...

Страница 4: ...ument is entrusted to the recipient for personal use only Any impermissible transfer duplication translation into other languages or excerpts from this operating manual are prohibited The manufacturer assumes no liability for print errors 1 3 Further information _ Software version of the product V1 1 or later _ For the exact product name refer to the type plate on the rear side of the product ...

Страница 5: ... product can only be guaranteed if applicable safety precautions and the device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury or death as well as property damage can occur DANGER Incorrect area of application In order to prevent erratic behaviour of the product personal injury and property damage the product must be used exclusively a...

Страница 6: ...1113 Differential pressure 200 hPa 2000 hPa Max resolution 0 1 Pa 1 Pa Usual applications include precise measurements of filter condition gas flow pressure draught leak integrity dynamic pressure flow speed The pressure connection is made at the supplied interchangeable pressure connection ports with suitable hoses The product must only be used under the conditions and for the purposes for which ...

Страница 7: ...elements _ Display _ Battery indicator Evaluation of the battery status _ Unit display Display of the units or Min Max Hold information text _ Main display Measurement of the current pressure or value for min max hold _ Auxiliary display Measurement of the current pressure in Min Max Hold mode _ Bar graph Trend display with the special function F NE ...

Страница 8: ...on the product Activate deactivate lighting Long press Switch off the product Reject changes in a menu _ Up Down button Press briefly Display of the min max value Change value of the selected parameter Long press Reset the min max value of the current measure ment Both simultaneously Rotate display overhead display _ ...

Страница 9: ...Long press 2s Start menu configuration ONF appears in the display Close menu changes are saved Long press 4s Depending on the selected special function Activa tion of the Tare function NVLL high resolution measurement F NE or rapid measurement with mean value AVR 3 4 Connections Universal connection Interchangeable pressure connection via G1 8 thread ...

Страница 10: ...pears in the display Release the Function key _ Parameter Values Meaning _ Display unit UN T Pa only at RS 1107 HPa mBaR BaR only at RS 1113 PSI mmKG _ Activatable special functions NVLL Tare function available F NE High resolution measurement with 0 1 Pa RS 1107 or 1 Pa RS 1113 activatable AVR 0 02 AVR 0 05 AVR 0 10 Rapid measurement with mean value over 2 s 5 s 10 s activatable _ ...

Страница 11: ...justed to the lowest value e g 20 00 20 00 hPa RS 1107 200 0 200 0 hPa RS 1113 _ Shut off time POFF OFF No automatic shut off 0 15 0 30 1 00 4 00 12 00 Automatic shut off after a selected time in hours and minutes during which no buttons have been pressed _ Backlight L TE OFF Backlight deactivated 0 15 0 30 1 00 4 00 Automatic shut off of the backlight after a selected time in minutes and seconds ...

Страница 12: ...product 4 Release the On Off button after 1 second and then the Down button in order to call up the adjustment menu The display shows the first parameter Parameter Values Meaning _ Zero point correction offset PR OF 0 00 No offset 5 00 5 00 Offset active in hPa at RS 1107 50 0 50 0 Offset active in hPa at RS 1113 Gradient correction slope PR SL 0 00 No slope 5 00 5 00 Slope active in Formula Zero ...

Страница 13: ...rpose 6x1 mm PVC Up to 5 bar rel vacuum suitable 6x1 mm PE Up to 10 bar rel vacuum suitable 6x1 mm PUR Up to 9 bar rel vacuum suitable 5 1 Special functions _ With the special functions that can be selected via the Configuration menu the device can be optimised for special measuring tasks After it is switched on the device starts up in standard measuring mode the relevant special function is start...

Страница 14: ... adjusted pressure applications In the Configuration menu the special function FVNC F NE has been selected The high resolution measurement can be activated by pressing and holding the Func tion key for 4 s Then the sensor is immediately zeroed and the optimised parameters for this measurement are activated CAUTION When starting the special function make sure that there is no pressure at the connec...

Страница 15: ... and the optimised parameters for this measurement are activated CAUTION When starting the special function make sure that there is no pressure at the connec tions NOTE The increased current consumption in this mode decreases battery life The quickly determined measurement replaces other devices such as a U tube ma nometer The four bars in the lower display provide additional support The two middl...

Страница 16: ...himney draft tests of forced air burners and consequently conventional electronic manometers are not adequate for task This special function optimises the device for this application purpose The different mean value times of 2 5 or 10 seconds can be selected depending on the requirement _ The first parameter is shown in the auxiliary display _ By pressing and holding the Function key for 2 s the s...

Страница 17: ...rs in the main display the battery voltage is no longer adequate for operation of the product The battery is fully depleted 6 2 2 Changing battery _ DANGER Danger of explosion Using damaged or unsuitable batteries can generate heat which can cause the batter ies to crack and possibly explode Only use high quality and suitable alkaline batteries _ CAUTION Damage If the batteries have different char...

Страница 18: ... screws A and remove the cover 2 Carefully replace the two Mignon AA batteries B En sure that the polarity is correct It must be possible to in sert the batteries in the correct position without using force 3 The O ring C must be undamaged clean and posi tioned at the intended depth In order to facilitate assem bly and avoid damage a suitable grease can be applied 4 Fit the cover on evenly The O r...

Страница 19: ...urement too high Product is defective Stay within allowable meas urement range Send in for repair ERR 2 Measuring range is undercut Measurement too low Product is defective Stay within allowable meas urement range Send in for repair ERR 3 Display range has been exceeded Incorrect display unit Incorrect resolution F NE Function active Correct setting Deactivate function ERR 4 Display range has been...

Страница 20: ... 1700 hPa Pressure connection 2 hose connections interchangeable with G1 8 universal ports Measuring cycle FAST approx 25 measurements per second SLO approx 2 5 measurements per second Display 3 line segment LCD additional symbols illuminat ed white duration adjustable Standard function Min Max Hold Auto power Off function if activated the product switches off automatically Activatable special fun...

Страница 21: ...attery life approx 1 mA slow measurement SLO Operating time approx 3000 h Battery indicator 4 stage battery status indicator Note for low battery voltage BAT LO Directives and standards The devices conform to the following Directives of the Council for the harmonisation of legal regula tions of the Member States 2014 30 EU EMC Directive 2011 65 EU RoHS Applied harmonised standards EN 61326 1 2013 ...

Страница 22: ...ure connection 2 hose connections interchangeable with G1 8 universal ports Measuring cycle FAST approx 25 measurements per second SLO approx 2 5 measurements per second Display 3 line segment LCD additional symbols illuminat ed adjustable white permanent illumination Standard function Min Max Hold Auto power Off function if activated switches the product off automatically Activatable special func...

Страница 23: ...t Battery life approx 1 mA slow measurement SLO Operating time approx 3000 h Battery indicator 4 stage battery status indicator Note for low charge level BAT LO Directives and standards The devices conform to the following Directives of the Council for the harmonisation of legal regula tions of the Member States 2014 30 EU EMC Directive 2011 65 EU RoHS Applied harmonised standards EN 61326 1 2013 ...

Страница 24: ...ating Manual RS 1107 RS 1113 rspro com 9 Service 9 1 Manufacturer If you have any questions please do not hesitate to contact us RS Components Limited Birchington Road Corby Northamptonshire NN17 9RS WEE GF0002ZR Contact ...

Отзывы: