![RS PRO 192-4382 Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/rs-pro/192-4382/192-4382_instruction-manual_1496053003.webp)
2
3
Sensor probe
F
N
Working principle
Magnetic induction
Eddy current principle
Measuring range
0~1350µm
0~1350µm
0~53.1mils
0~53.1mils
Guaranteed tolerance
(of reading)
0~1000um ±(2.5%+2um)
0~1000um ±(2.5%+2um)
1000~1350um(±3.5%)
1000~1350um (±3.5%)
0~39.3mils (±2%±0.08 mils) 0~39.3mils (±2%±0.08mils)
39.3~53.1mils(±3.5)
39.3~53.1mils(±3.5)
Precision
0~100um(0.1um)
0~100um(0.1um)
100um~1000um(1um)
100um~1000um(1um)
1000um~1350um (0.01mm)
1000um~1350um (0.01mm)
0~10mils(0.01mils)
0~10mils(0.01mils)
10~53.1mils(0.1mils)
10~53.1mils(0.1mils)
Minimum curvature
radius
1.5mm
3mm
Diameter of Minimum
area
7mm
5mm
Minimum measurable
thickness
0.5mm
0.3mm
Overload display
----
----
Working temperature
0°C~40°C(32°F~104°F)
Working relative humidity 20%~90%
3. Specifications
2.3 Supply Schedule
• Gauge with two AAA 1.5V batteries, plastics carrying case, operating instructions (English),
steel and aluminum substrate, and standard flake for calibration.
2.4 Probe
The Probe systems are spring-mounted in the probe sleeve. This ensures safe and stable
positioning of the probe and constant contact pressure.
A V-groove in the sleeve of the probes facilitates reliable readings on small cylindrical
parts. The hemispherical tip of the probe is made of hard and durable material. Hold the probe
by the spring mounted sleeve and put on measuring object.
Coating thickness tester / English
Coating thickness tester / English
3.1 Front View
1- LCD display
2- Select key: For ON/OFF, select and ok
3- Up key: For upward selection and
working mode switch
4- Clear Key: Calibration Zone of the FE
and NFE
5- Back Key: For return or delete the
last measured data
6- Down key: For downward selection or
working mode switch
7- Battery cover
3.2 LCD Display
1- Working mode
2- Measured data quantity in the working mode
3- Auto measure mode
4- Ferrous or No-ferrous metals being measured.
It will change automatically in the Auto measure mode
5- Battery indication
6- Measure data
7- Unit
8- Delete data, press Back button to delete the current data
4. Ready to start
4.1 Power Supply
For power on, please press and hold button. If no LCD display: please check if the battery is
correctly installed or it is in low battery.
Display
: Please replace battery immediately.
Note that the gauge will make faulty measurements if the voltage is very low.
Note:
When power on, please away from any metal at least 10 cm above Before each
measurement, zero calibration is recommended.