10.5 - Reagents replacement
The Interrliner is equipped with the genuine Mechatronics bulk reagent containers with level sensors. Each
container has a specific position, see Reagent location label.
To replace the bulk reagent containers:
1. Open cabinet.
2. Remove level sensors and spacers from the empty container.
3. Remove the container screw caps and pull the necks of the bottle packs out of the cardboard box.
4. Install the level sensors and spacers.
5. Add new reagent information in the software.
WARNING:
Wrongly placed pickup tubes causes incorrect results or instrument malfunction.
Make sure to place the appropriate level sensors in the containers by checking the colored
labels/colored tubes matching the color codes on the containers.
Maintenance
74
MRN-164_2-EN Version 1.08 Starrsed TL
Содержание STARRSED TL
Страница 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Страница 2: ......
Страница 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Страница 13: ...Graphic showing typical absorption values of a sample Introduction MRN 164_2 EN Version 1 08 Starrsed TL 13...
Страница 105: ...7 8 9 10 11 12 13 Clean liquid separator Removing Work instruction Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 105...
Страница 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...
Страница 121: ...13 Appendix for Interrliner Appendix section Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 121...
Страница 125: ...TLAX079200 STLAX079200 Tubing set Starrsed TL Appendix for Interrliner MRN 164_2 EN Version 1 08 Starrsed TL 125...