2
Instrument description
The
Starrsed RS
can accommodate sample racks from different manufacturers. The individual rack with the
blood samples is snapped into the RR Mechatronics rack adapter and need only be placed on the entry
platform of the Starrsed Rack unit.
The operator simply presses the button "Sample mode".
The instrument picks up the rack, barcode labels are read and the blood is mixed by rotating the rack 8
times, as recommended by the ICSH. Blood is aspirated from the first sample tube. Thereafter the rack is
rotated one time to ensure each blood sample is thoroughly mixed and the sample is picked up to be
aspirated. Aspiration takes place via Mechatronics proprietary double needle mechanism.
The citrate dilution takes place in a 4+1 ratio and is achieved with ± 3% accuracy.
Eighty-four Westergren pipettes are housed in the carousel. Each is of precision bore glass. After each cycle,
the pipette is cleaned automatically with low foam detergent followed by a drying cycle.
The fill line is back-flushed using saline solution.
The temperature is corrected to the standard value of 18°C and ESR's may be read after one hour or 30
minutes. A foretell one-hour result is presented in the 30-minute mode.
The
Starrsed RS
can be interfaced bidirectionally with Laboratory Information Management Systems (LIMS)
through a variety of interface protocols.
Results of the test are expressed in millimeters. This data together with the patient ID number is sent to the
Laboratory Information System along with the sedimentation time used (60 or 30 minutes), the temperature
and the dilution ratio.
Blanco
Instrument description
MRN-071-2-EN version 7.02 Starrsed RS
23
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...