1. The gripper moves to the needle position.
2. The needle comes down.
3. Fill a container with 150 ml hot (80 °C) de-ionized water.
4. Add 15 ml cleaning agent. (QRR 010905)
5. Stir the prepared solution.
6. Put the container close to the needle location.
7. Push the silicon tube over the sample probe.
8. Press
Continue
.
9. The fill and clean process is started.
13.4 - Level 3 maintenance
Level 3 maintenance is level 4 maintenance adding the following extra's.
1. Replace the pinch valve tube
ESRI010246.
2. Replace the blue disk filter
QWLV040003.
3. Replace the Peristaltic waste pump cassette
ESRI090921
including the Blotting washer
ESRI090026.
Be careful, as there may be blood in the cassette. First, prepare some disinfectant and put this in the
waste separator. Press
Prime Disinfectant
to pump disinfectant through the pump cassette.
Symptoms for a bad or faulty waste pump cassette:
▪
Waste separator error.
▪
Taking too long before the separator empties.
Detailed instructions of this procedure can be found in the Work Instruction
.
13.5 - Level 2 maintenance
Level 2 maintenance contains all maintenance actions from level 3 maintenance and the following:
Replace the Teflon tip (from repair set QWLV030901) on the syringe of the diluter assembly. (Not needed if
Syringe QWLV030021 is installed).
See Dis- and re-assembly diluter syringe
General maintenance
136
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...