Waste disposal
The waste container has a level sensor and as soon as the level sensor generates a waste error, the waste
container must be emptied. The waste must be treated as potentially infectious (biohazardous) material and
disposed of according to local regulations. Preferably, discard the complete waste container and replace it
with a cleaned one. Clear the error.
If the waste line is to be connected to a centralised waste collection system, the following requirements
must be met:
1. Waste tube must not exceed 5 meters or 18 feet in length.
2. Drain height must not be higher than the original waste container inside the instrument.
Disclaimer:
Check the specifications of the central waste system for rules about discharging the
waste.
9.1 - Replacing the waste container
1. Lift the left cover and pull the waste container forwards.
2. Unscrew the cap.
3. Place the new waste container and tighten the screw cap.
4. Lift the left cover and place the waste container back into the ESR analyser unit.
Note
: If you are re-cycling waste containers, make sure that they are bleached and rinsed thoroughly.
Waste disposal
MRN-164_2-EN Version 2.02 Interrliner