The Clear rack history button will clear the contents of the rack history file and restart to build-up a new rack
history file.
5.3.4 - ESR Statistics screens
A statistical graph is produced over a selected period. Make a selection of the following graphs;
▪
Daily mean (mm)
Use this to check variations in the daily mean ESR.
▪
Hazy (%)
Increasing hazy aspects are an indication for contamination of the instrument, see
▪
Dilution (%)
Increasing dilution errors indicate the need for maintenance of the diluter system.
▪
Bubbles on top (%)
Increasing samples with bubbles indicate the need for maintenance of the aspiration system, see
▪
ESR errors (%)
Increasing ESR errors may indicate the need for maintenance, see
▪
Number of samples
This can be used to document variations in work load.
5.3.5 - QC Results screens
In this section results and statistics from QC samples are shown, in the section
links can be
created between QC sample ID's and Lab ID's.
The results from Starrsed Control
are separated on their own tabs. Both tabs have the
same layout and options. Results can be displayed in table format or in graphical format.
Starrsed program
36
MRN-164_2-EN Version 2.02 Interrliner