9. 

Ethernet local area network 

ROTRONIC devices with an Ethernet (TCP/IP) interface presently use an internal module 
manufactured by Digi International (Digi Connect ME for a wired connection, Digi Connect WI-ME 
for a wireless connection). Both types of module feature a web interface that is used to configure 
the module.  
 

IMPORTANT:

 The TCP/IP settings of both the HygroClip DI-4 and DI-4-WL should be configured 

to be compatible with the local area network. Detailed instructions are provided separately in 
document 

IN-E-TCPIP-Conf_10.doc

 (also available in PDF format).  

 
Both the HygroClip DI-4 and DI-4-WL are shipped with a Device Configuration Certificate that 
provides information about the factory configuration settings of both the Ethernet module and 
device. See example below: 
 

Setting 

Value 

DHCP or Static Address 

Static 

IP Address 

192.168.1.1 

Subnet Mask 

255.255.255.0 

Default Gateway 

192.168.1.0 

TCP Port 

2101 

MAC Address 

00:40:9D:28:2D:2A 

Firmware Release (device) 

1.4 

User Name (web interface) 

rotronic 

Password (web interface) 

wlan 

Serial Port Configuration Profile 

TCP Sockets 

Basic Serial Settings 

 

 Baud 

Rate 

57600 

 Data 

Bits 

 Parity 

None 

 Stop 

Bits 

 Flow 

Control 

None 

 

10.  RS-485 multi-drop network 

 
All models of HygroClip DI feature a RS-485 connector. When the number of available PC ports 
is limited, this connector can be used to connect together up to 64 HygroClip DI (or other devices) 
in a multi-dropped arrangement. In principle, an unlimited number of such networks can be 
monitored with the HW4 software, but each RS-485 multi-drop network is limited to 64 devices. 
The HygroClip DI can be used either as a slave or a master, without special configuration. The 
master is automatically the device that is directly connected to the PC or LAN by means of a USB 
port or TCP/IP port. 
 

RS-485 Address

: the RS-485 address is used to identify the individual devices that are 

connected to a RS-485 multi-drop and must be unique to each device. This address is part of the 
HygroClip DI communications protocol. The factory default for the RS-485 address is 00. In 
general, this address should not be changed manually (see note below). As far as HW4 is 
concerned, all masters can share the same RS-485 address. When the HW4 command “Search 

Содержание HygroClip DI tal interface

Страница 1: ...rf Switzerland Document code Unit Instruction Manual Document Type HygroClip DI digital interface instruction manual Document title Page 1 of 24 2007 Rotronic AG IN E HyClip DI V1_10 doc HygroClip DI...

Страница 2: ...9 Ethernet local area network 18 10 RS 485 multi drop network 18 11 Specifications 20 12 Accessories 21 13 Communications protocol 21 14 Electrical Installation Guidelines 22 15 Document releases 24...

Страница 3: ...r than the capacity of the PC there is no limit to the number of HygroClip DI that can be networked with the PC The main functions of the HW4 software include on line data monitoring and recording by...

Страница 4: ...ted 12 VDC 100 240 VAC adapter 170 mA 4 Probes The HygroClip DI is designed for use with the ROTRONIC HygroClip digital humidity temperature probes These are highly accurate probes which are adjusted...

Страница 5: ...in air max 85 C 185 F wire mesh filter HygroClip SP05 measurement in air ducts max 85 C 185 F HygroClip SC05 measurement in tight spaces max 100 C 212 F HygroClip HP28 insertion probe measurement of...

Страница 6: ...le model AC1616 is not compatible with the HygroClip DI HygroClip IC 1 100mm IC 3 250mm for through wall installation max 200 C 392 F wire mesh filter HygroClip IM 1 120mm IM 3 270mm for through wall...

Страница 7: ...Note use pins 3 and 4 to power multi dropped instruments with a single external 12 VDC power supply with adequate mA rating Pin 4 and the 12VDC receptacle normally used for the AC adapter are not conn...

Страница 8: ...security When new software is being installed the NET framework also eliminates the potential problem of conflicting dynamic library files DLL According to Microsoft the NET framework will be used by...

Страница 9: ...e manager menu bar is located at the top of the form File Open opens the device configuration directory specified in the HW4 Settings Form File Locations Tab and displays all available device configur...

Страница 10: ...mation As far as possible use a unique device name RS485 address click on the underlined link to change the instrument address to be used in conjunction with an RS 485 network multi drop Each network...

Страница 11: ...orm to globally change the unit system used by the instrument If the unit system offers several choices such as In Hg or PSI for pressure choose a unit by left clicking on the arrow to the right of ea...

Страница 12: ...Input Name use a maximum of 12 characters Calculation Each probe input of the HygroClip DI can be set to compute one of the following psychrometric parameters Dew point Dp Frost point Fp Wet bulb tem...

Страница 13: ...Alarm alarm conditions can be defined for humidity temperature and the calculated parameter Values that are below the low alarm value or above the high alarm value will trigger an alarm A value can b...

Страница 14: ...ctronic signatures HW4 is designed to keep track of configuration changes and of any probe adjustment HW4 also keeps track of user events main operations performed by each user and self events softwar...

Страница 15: ...al interface instruction manual Document title Page 15 of 24 2007 Rotronic AG IN E HyClip DI V1_10 doc To protect the device click on the underlined link HW4 opens the following form where a password...

Страница 16: ...the underlined link HW4 opens the password form Enter the password and click on the OK button HW4 confirms that the device is now unprotected FORGOT THE PASSWORD should you forget the password used t...

Страница 17: ...M33 transmitters HygroClip Alarm programmable alarm card HygroClip DI interface HygroStat MB Thermo Hygrostat IMPORTANT All devices connected to a RS 485 multi drop network master and slaves must use...

Страница 18: ...0 TCP Port 2101 MAC Address 00 40 9D 28 2D 2A Firmware Release device 1 4 User Name web interface rotronic Password web interface wlan Serial Port Configuration Profile TCP Sockets Basic Serial Setti...

Страница 19: ...p DI with USB port that can be directly connected to a PC depends on the number of USB ports available on the PC PC HW4 USB port Master Address 1 Address 2 Address 64 DI USB RS485 RS485 RS485 Slave Sl...

Страница 20: ...00 Data bits 8 Stop bits 1 Parity none Probe inputs 2 HygroClip humidity temperature probes only Operating limits electronics 0 100 RH 40 85 C 40 185 F Measuring range 0 100 RH HygroClip probes are av...

Страница 21: ...length 2 meter 6 5 ft or 5 meter 16 4 ft T7 03 B5 Adapter cable for HygroClip probes with T7 connector Cable length 3 meter 9 8 ft AC1211 AC adapter 12VDC 170mA includes adapter for US receptacles 2 1...

Страница 22: ...s are tested for Electromagnetic Compatibility according to following European standards EN 61000 6 3 EN 61000 6 1 residential EN 61000 6 4 EN 61000 6 2 industrial Whenever the level of electromagneti...

Страница 23: ...al the RS 485 cable should be shielded and comply with the following specifications Cable capacitance 300pF m or 90 pF ft Line impedance 100 15 Line resistance 140 km or 225 Signal lines Twisted pair...

Страница 24: ...ubes or supporting constructions should be integrated into the system Protect the potential equalization cable and connections against corrosion Select the cross section of the potential equalization...

Отзывы: