11. Troubleshooting
Symptom
Possible Reason
Solution
Display unit does not turn
on
Battery is discharged
Recharge battery
Verify battery is in the battery
compartment
Place battery in the battery
compartment
Faulty electronics
Contact ROTHENBERGER
Service Center
Display turns on, but does
not show image.
Loose cable connection
Check cable connections.
Imager head covered by debris
Clean imager head
Display shows an
unresponsive image
Image processor encountered
an error
Turn OFF unit/turn ON to
reset processor
Battery does not hold a
charge
Battery may have discharged
slightly during storage
Remove battery when
storing
Run time dependent on LED
usage
Recharge battery more often
when extensive use of LEDs
Battery may be damaged
Replace battery
Life cycles of battery expired
Replace battery
No Signal
Wrong frequency
Verify frequency of your
sonde and adapt ROLOC
frequency via menu
Depth measurement shows: ROLOC PLUS is out of range
Verify strongest signal
ROSCOPE 1000/i2000
Error Code 1
“unplugged”
Error present when imager has
been plugged in and working,
and then has been removed
Contact Service Department
ROSCOPE 1000/i2000
Error Code 2
“Digital / No Video Sync”
Indicates that a digital imager has
been identified but no video sync
are present.
Contact Service Department
ROSCOPE 1000/i2000
Error Code 3
“Analog / No Video Sync”
Indicates that an analog imager
has been identified but no video
syns are present.
Contact Service Department
ROSCOPE 1000/i2000
Error Code 4 (PPI)
“Video Data Not in Correct
Format”
Indicate that the video data is not
in the correct format.
Contact Service Department
ROSCOPE 1000/i2000
Error Code 5
“Aurora Detected”
This is a condition where the
imagers automatic correction
scheme (digital imager only)
results in an image with video
artifacts – the imager is then
reset. If this condition persists
after subsequent retries, this
error is then presented.
Contact Service Department