9.  Do not soak more than 3 cm from the tip of the sampling stick (comb) into water or other liquid 

samples. Water or other liquid samples which wet the shaft of the sampling stick (green) may 
affect test results. 

10.  Ensure that the LuciPac

®

 A3 Aqua is not shifted or displaced, discard the item, because 

accuracy cannot be guaranteed, swabs are not removable from the Lumitester PD30/Smart,  
or the instrument may cause an error (malfunction). 

11.  If the test reagent is leaking please don't use the swab! 

 

7.  Measurement 

 

Complete the procedures listed below within a temperature range of 20 °C to 35 °C (68 °F to 
95 °F). Make sure to always run measurement tests at the same temperature to maintain 
repeatability for comparison. Remove the LuciPac

®

 A3 Aqua from refrigerator and wait until they 

have reached room temperature (about 20 min, 20 °C to 35 °C (68 °F to 95 °F)). Use the sampling 
devices then as soon as possible. 
1.  Remove the sampling stick from the main body (casing). 
2.  Soak sampling stick about 3 cm from the tip of the sampling stick into water or other liquid 

samples, and shake gently. Make sure no bubbles remain within sampling stick's comb. 

3.  Pull out the sampling stick slowly and straight up from the sample. 
4.  Return the sampling stick to the main body (casing) and push it all the way into the main body. 
5.  Push firmly onto LuciPac

®

 A3 Aqua casing and agitate. 

6.  Allow the leftover luminescent reagent to thoroughly dissolve. 
7.  Insert the LuciPac

®

 A3 Aqua into the Lumitester to measure the results. Press ENTER. 

8. 

Remove the LuciPac

®

 A3 Aqua case from the measuring chamber.

 

 

8.  Safety recommendations 

 

Please observe the following items to ensure safe use of this product.

 

 

Please note safety recommendations given in the Safety Data Sheet! 

 

Do not attempt to drink the kit reagent or touch it with bare hands or allow it to splash into 
eyes. Please make sure to read the precautions and instructions in this Instruction Manual 
before attempting to use the kit and exercise extreme caution when using it. 

 

Be careful not to get fingers caught when inserting the sampling stick into the main body 
(casing). 

 

Handle with care when storing and disposing of the device and its reagents to ensure that 
none of the substances become mixed with food and other products. 

 

Please make sure to store this kit and its parts out of the reach of young children. 

 

Take precaution when disposing of this kit after use to ensure that such substances do not 
become mixed with food products at food production centers and similar facilities 

 
 
 

9.  Short instruction 

Prior to use, make sure the LuciPac

®

 A3 Aqua sticks have come to room temperature.  

But then use immediately. 

 

Содержание LuciPac A3 Aqua

Страница 1: ...tored at a low temperature 2 C to 8 C 35 6 F to 46 4 F for long term storage The kit can be stored below 25 C 77 F for up to 14 days or below 30 C 86 F for up to 5 days before opening an aluminum bag without any adverse effect on the long term stability Do not freeze the device We recommend that you use all 20 sampling devices in a single bag at one time after opening an aluminum bag If you have l...

Страница 2: ...rase pyruvate orthophosphate dikinase PPDK and a pyruvate kinase PK The method produces a defined amount of luminescence proportional to the amount of adenosine triphosphate ATP adenosine diphosphate ADP and adenosine monophosphate AMP in a sample Firefly luciferase emits light in the presence of ATP and luciferin The AMP generated from this reaction is converted back to ATP using the PPDK while e...

Страница 3: ...ted environment 2 C to 8 C 35 6 F to 46 4 F High temperatures may cause product performance to drop 5 Do not subject the device or any part of it to direct sunlight for long periods of time Strong light may cause product performance to drop 6 Do not touch any of the parts inside the device particularly not any part of the sampling stick comb itself with a finger or other object before use Touching...

Страница 4: ... sampling stick to the main body casing and push it all the way into the main body 5 Push firmly onto LuciPac A3 Aqua casing and agitate 6 Allow the leftover luminescent reagent to thoroughly dissolve 7 Insert the LuciPac A3 Aqua into the Lumitester to measure the results Press ENTER 8 Remove the LuciPac A3 Aqua case from the measuring chamber 8 Safety recommendations Please observe the following ...

Страница 5: ...606 0 Fax 49 0 721 5606 149 info carlroth com www carlroth com The company is a limited partnership with headquarters in Karlsruhe reg court Mannheim HRA 100055 Roth Chemie GmbH with headquarters in Karlsruhe reg court Mannheim HRB 100428 is the personally liable partner Managing Director André Houdelet Sales tax identification number DE 143621073 sse 07 2021 ...

Отзывы: