Rossmax MANDAUS II Скачать руководство пользователя страница 2

3.  With  the  valve  closed, 

press the bulb and con-
tinue pumping to a val-
ue  30-40  mmHg  above 
your normal blood pres-
sure.  (Fig. 3)

4.  

Open the valve to de-
flate the cuff gradually 
at  a  rate  of  2-3  mmHg 
per second. (The defla-
tion rate is displayed on 
LCD.) (Fig. 4)

5.  Record the onset of Ko-

rotkoff sound as the sys-
tolic pressure, and the 
disappearance of these sounds as diastolic pressure.

6.  After  measurement  is  completed,  fully  open  the  valve  to 

release any remaining air in the cuff. The heartbeat rate is 
displayed on the LCD. (Fig. 5)

4.  Using the correct cuff size is important for an accurate 

reading.

5.  Display Explanations

7.  Applying the Cuff

3.  Blood Pressure Standard

1.  Introduction

4.  Name/Function of Each Part

2.  Preliminary Remarks

Attention:

  

Consult the accompanying documents. 
Please read this manual carefully before 
use. Please be sure to keep this manual. 

Mandaus II complies with the European regulations and bears 
the  CE  mark  "CE  0366".  The  quality  of  the  device  has  been 
verified  and  conforms  to  the  provisions  of  the  EC  council  di-
rective 93/42/EEC (Medical Device Directive), Annex I essential 
requirements and applied harmonized standards.

EN 1060-1: 1995/A1: 2002 Non-invasive sphygmomanometers 
- Part 1 - General requirements 

EN 1060-3: 1997/A2: 2009 Non-invasive sphygmomanometers 
-  Part  3  -  Supplementary  requirements  for  electro-mechanical 
blood pressure measuring systems

8.  Measurement Procedures

1.  Place the stethoscope head over the 

main artery, underneath the artery 
mark of the cuff. (Fig. 1)

2.  

Press the “ ” key. After zeroing, 
Mandaus II is ready to measure. (Fig. 
2) 

Thank you for choosing Rossmax Mandaus II, the shock and 
splash resistant backlighted digital sphygmomanometer. Blood 
pressure measurements determined with Mandaus II are equiv-
alent to those obtained by a trained observer using cuff/stetho-
scope auscultation method. 

Mandaus II is protected against manufacturing defects by an 
established International Warranty Program. For warranty infor-
mation, you can contact the manufacturer, Rossmax Interna-
tional Ltd. or your local distributors. 

The 

National High Blood Pressure Education Program 

Coordinating Committee

 has developed a blood pressure 

standard, classifying blood pressure ranges into 4 stages. 

(Ref. The Seventh Report of the Joint National Committee on Prevention, De-
tection, Evaluation, and Treatment of High Blood Pressure-Complete Report 
JNC-7, 2004)

This blood pressure classification are based on historical data, 
and may not be directly applicable to any particular patient.
It is important that you consult with your physician regularly.  
Your physician will tell you your 
normal blood pressure range 
as well as the point at which 
you will be considered at risk 
before making any healthcare 
decision based on the device's 
readings.
For  reliable  monitoring  and 
reference of blood pressure, 
keeping long- term records is 
recommended. Please down-
load the blood pressure log 
(chart for recording and track-
ing crucial info: date, time, 
blood pressure, pulse rate and 
body condition) at www.ross-
maxhealth.com.

Arm Cuff

Air Tube and 

Connector

Inflating 

Bulb

Air  

Release 

Valve

Back light key

On/Off/Zeroing key

Carrying Case

Stethoscope

Index Line

Artery Mark

Arm Circumference Indicator

1.  Blood Pressure

2.  Deflation Rate (During Measurement)/  

Heartbeat Rate (After Measurement)

3.  Heartbeat Mark

4.  Weak Battery Mark

5.  Error Mark

6.  Calibration Reminder

1

2

4

3

5

6

1

2

4

3

5

6

6.  Installing and Replacing the Batteries

1.  Rotate the clip through 90 °. 

2.  Loosen the screw on the battery cover. 

3.  Install the batteries with correct polarities. 

4.  Replace the battery cover and tighten the screw. 

5.  Rotate the clip back.

You need to replace the batteries when:

1. low battery icon appears on display.
2.   the  On/Off/Zeroing key is pressed and nothing appears on 

display.

Caution:

 1.  Batteries are hazardous waste. Do not dispose them 

together with the household garbage.

               2.  There are no user serviceable parts inside. Batteries 

or damage from old batteries are not covered by 
warranty.

               3.  Use exclusively brand batteries. Always replace with 

new batteries together. Use batteries of the same 
brand and same type. 

1

2

4

3

5

6

1.  Plug the connector into the gauge 

and twist clockwise to secure the 
connection.

2.  Place the cuff over the bare upper 

arm, wrap it with the tube point-
ing your palm, and the artery mark 
over your main artery. 

Applicable cuff circumferences

Rossmax Gauge Cuffs

Young Adult (S)

18 - 26 cm (7.0”-10.3”)

Adult (M)

24 - 36 cm (9.5” – 14.2”)

Large Adult (L)

34 - 46 cm (13.4”-18.1”)

3.  The  edge  of  the  cuff 

should be at approximate-
ly 1.5 to 2.5 cm above the 
inner side of the elbow 
joint. If the index line falls 
within the range of the 
arm circumference indica-
tor, the cuff circumference 
is suitable, otherwise you 
may need a cuff with a dif-
ferent circumference. 

5.  To  stop  operation  at  any  time,  open  the  air  release  valve, 

and the air in the cuff will be rapidly exhausted. 

1

2

4

3

5

6

1

2

4

3

5

6

Fig. 1

Fig. 2

Fig. 3

Fig. 4

1

2

4

3

5

6

1

2

4

3

5

6

Fig. 5

1

2

4

3

5

6

1

2

4

3

5

6