User Manual
RPC-N 75
Ω
Calibration Kit
P5CK14A-170 and P5CK010-170
P5CK14A-170
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.com
Tel. : +49 8684 18-0
Email : [email protected]
Page
4 / 18
D
ie
s
e
s
D
o
k
u
m
e
n
t
is
t
u
rh
e
b
e
rr
e
c
h
tl
ic
h
g
e
s
c
h
ü
tz
t
●
T
h
is
d
o
c
u
m
e
n
t
is
p
ro
te
c
te
d
b
y
c
o
p
y
ri
g
h
t
●
R
o
s
e
n
b
e
rg
e
r
H
o
c
h
fr
e
q
u
e
n
z
te
c
h
n
ik
G
m
b
H
&
C
o
.
K
G
1.2 Warnings
The calibration standards and the gauges of this calibration kit are mechanically sensitive devices.
High accelerations e.g. when falling down will damage them.
Connect the calibration standards only to test ports and other devices that have been cleaned and
checked for gauge, concentricity and integrity of the contact fingers. Detailed procedures are
described in Chapter 6 and 7.
Tighten the connector coupling nuts to the recommended torque with the torque wrench. Do not
over torque!
Consider the power applied to the Load calibration standards. Maximum power limits for the sum of
RF and DC power are defined in the specific data sheets.
It is good lab practice to use at least an ESD wrist strap when handling calibration standards or
DUTs on a VNA although the calibration standards themselves are not ESD sensitive. Please
consult your VNA manual.
Follow further limitations given in the specific data sheets.
Never connect RPC-N 75
Ω
connectors to RPC-N 50
Ω
or other type N 50
Ω
connectors.
Specially the RPC-N 75
Ω
female center conductor will be be destroyed by an 50
Ω
male
center conductor.