The male and female Short standards of this kit are designed with the same offset length. 
So the TRM calibration technique can be used. It requires high reflects with a symmetrical 
reflection during  VNA calibration on both test ports. TRM needs less connections during 
calibration than OSLT. 
 

Recommendation

:  

The 

TRM

  calibration  is  the  recommended  calibration  method  to  use  as two-port  calibra-

tion. It can be used in a SMP male-male, female-female and male-female test port config-
uration.  
The standard definition files for the VNAs include a special ‘Reflect’ device, please use it 
for a TRM calibration. If no file for the specific VNA is available the standard definitions for 
all calibration standards have to be entered manually. The ‘Reflect’ device is derived from 
the short calibration standard but the L0, L1, L2 and L3 are all set to zero. This keeps the 
symmetry of the ‘Reflect’ that is necessary for the TRM calibration. 

 
 

Calibration Procedure 

 
The procedures to perform a VNA calibration are very different between the VNA models. Mod-
ern VNA often guide you thru the settings and actions. For details please consult the manual of 
your specific VNA. 
 

 

4.2 Calibration Check 

 

It  is  advisable  to  check  the  calibration  after  error  correction  has  been  performed,  particularly 
when test port cables and adapters have been used. 
 
During the calibration 

  Calibration standards can easily be mixed up 

  Connections may not be tightened or aligned correctly 

 
There might be defects 

  VNA, e.g. mechanically overloaded test port 

  Instable test port cables 

  Mechanically overloaded load 

  Bent or broken contact fingers 

 
There is no Verification Kit available for SMP connectors. So additional loads and adpaters can 
be used to verify the quality of reflection and transmission measurements or short cable assem-
blies  of  high  quality may  be  used to  check for  plausible  measurement  results.  This  should  be 
done from time to time to identify defects.  
 

 

 

Содержание 19CK010-150

Страница 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 16 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Страница 2: ...Kit Documentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and Mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installat...

Страница 3: ...ments of components with SMP connectors RPC 2 92 to SMP adapters are included in this calibration kit to interface with standard 2 92 mm K or 3 5 mm test port cables and VNAs 1 2 Warnings The calibrat...

Страница 4: ...plug plug 19S121 S20D3 1 Calibration adaptor jack jack 19K121 K20D3 1 Calibration adaptor plug jack 19S121 K20D3 1 Calibration adaptor plug plug 02S119 S20D3 1 Calibration adaptor jack jack 02K119 K20...

Страница 5: ...ing temperature range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined fo...

Страница 6: ...lex process Please consult the VNA operating man ual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously loading the...

Страница 7: ...ers Sweep type linear logarithmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Swe...

Страница 8: ...calibration standards have to be entered manually The Reflect device is derived from the short calibration standard but the L0 L1 L2 and L3 are all set to zero This keeps the symmetry of the Reflect...

Страница 9: ...case of re calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following se...

Страница 10: ...that have dents on the mating surfaces will usually also have raised edg es around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in ot...

Страница 11: ...irectly on to the con nector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connector to co...

Страница 12: ...destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may result i...

Страница 13: ...6 3 Handling and Storage Install the protective end caps and store the calibration devices in the foam lined storage case when not in use Never store connectors loose in a box or in a desk or bench dr...

Страница 14: ...the measurements and even make them unusable Adapting Test Ports The calibration kit comes with two RPC 2 92 to SMP adapters to adapt the test port of the VNA or the 2 92 mm test port cable to SMP in...

Страница 15: ...one or more contact la mellae which in turn can result in different measurement results This mated connection is kept free of me chanical stress This mated connection is bent too much and results in...

Страница 16: ...mall but the full detent version needs a high force to disconnect Care should be taken not to get hurt during this process and to keep the test port cable and the calibra tion standard under control T...

Отзывы: