Hazardous Area Oxymitter 5000
C-6
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
Table C-5. Channel 1 Status
The status of channel 1 is affected by the state of the unit alarm, as shown in Table C-5. In all cases, the channel
will read what it believes is the correct oxygen value. Self-clearing alarms are reset when the alarm condition goes
away. All others require the Oxymitter 5000 to be restarted
Alarm Condition
Channel 1 Status
Self-clearing?
Open Thermocouple
Bad
No
Shorted Thermocouple
Bad
No
Reversed Thermocouple
Bad
No
Heater Open Circuit
Bad
No
High High Heater Temperature
Bad
No
High Case Temperature
Bad
Yes
Low Heater Temperature
Bad
Yes
High Heater Temperature
Bad
Yes
Open Cell Circuit
Bad
Yes
High AC Impedance / Cell Bad
Uncertain
Yes
EEPROM Parameters Corrupt
Bad
No
Calibration Recommended
Good
No
Invalid Slope Value
Uncertain
No
Invalid Cell Constant Value
Uncertain
No
Bad Calibration
Uncertain
No
In Calibration
Uncertain
Yes
During Warm Up
Bad
Yes
Содержание Oxymitter 5000
Страница 2: ......
Страница 6: ......
Страница 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Страница 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...