![Rosemount 8712 Скачать руководство пользователя страница 169](http://html1.mh-extra.com/html/rosemount/8712/8712_reference-manual_1487460169.webp)
Reference Manual
00809-0100-4664, Rev BA
January 2010
E-15
Rosemount 8712
FISCHER AND PORTER
SENSORS
Connect coil drive and electrode cables as shown in Figure E-13.
Model 10D1418 Sensor
to Rosemount 8712
Transmitter
Figure E-13. Wiring Diagram for
Fischer and Porter Sensor
Model 10D1418 and Rosemount
8712
Table E-11. Fischer and Porter Model 10D1418 Sensor Wiring Connections
Rosemount 8712
Fischer and Porter
Model 10D1418 Sensors
1
L1
2
L2
Chassis Ground
17
3
18
1
19
2
ROSEMOUNT
8712
TRANSMITTER
Electrode Connections
U1
U2
G
8
6
7
L1
L2
5
3
2
1
Coil Connections
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.
Содержание 8712
Страница 2: ......
Страница 4: ......
Страница 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Страница 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Страница 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Страница 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Страница 131: ...Reference Manual 00809 0100 4664 Rev BA January 2010 B 7 Rosemount 8712 Figure B 1 CSA Installation ...
Страница 132: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 B 8 Figure B 2 Factory Mutual Hazardous Locations ...
Страница 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Страница 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Страница 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Страница 211: ......