70
Configuration, Startup and Operation
Section 3: Configuration, Startup and Operation
Reference Manual
January 2019
00809-0100-4892
ENTER when ready to start.
7.
Follow the prompts on the 6888Xi display to perform the trim.
8.
Use the RIGHT and LEFT arrow keys to select the digit to edit, then the UP and DOWN arrow
keys to change the value of the digit. Press ENTER to save the new value. Press ENTER to
accept the change.
9.
When complete, press EXIT to return to the ANALOG OUTPUT menu.
10. Repeat steps 1 through 9 for PROBE 2 if configured for dual channel.
D/A Trim with 6888Xi and Field Communicator
1.
Use the 375/475 Field Communicator to access the main HART menu.
2.
From the main menu, select SERVICE TOOLS.
3.
From the SERVICE TOOLS menu, select MAINTENANCE.
4.
From the MAINTENANCE menu, select ANALOG OUTPUT.
5.
From the ANALOG OUTPUT menu, select TRIM mA OUTPUT.
6.
In the first screen, a “Loop should be removed from automatic control” warning occurs.
Remove the 6888Xi from any automatic controls loops to avoid a potentially dangerous
operating condition. Press OK when ready.
7.
Follow the prompts on the 6888Xi display to perform the trim.
8.
Use the stylus on the screen to enter the new measured analog output value then press
ENTER to accept.
9.
When trim is complete, press OK to return to the ANALOG OUTPUT menu.
Содержание 6888Xi
Страница 2: ...B Table of Contents ...
Страница 4: ...II Table of Contents ...
Страница 8: ...VI Table of Contents Table of Contents Reference Manual January 2019 00809 0100 4892 ...
Страница 32: ...24 Installation Section 2 Installation Reference Manual January 2019 00809 0100 4892 ...
Страница 86: ...78 Troubleshooting Section 4 Troubleshooting Reference Manual January 2019 00809 0100 4892 ...
Страница 102: ...94 Replacement Parts Section 6 Replacement Parts Reference Manual January 2019 00809 0100 4892 ...
Страница 112: ...Appendix E 104 Appendix E China RoHS Table Reference Manual January 2019 00809 0100 4892 Appendix E China RoHS Table ...
Страница 113: ...105 ...