Errors
Error messages that may be displayed on the Integral Display, on a 275/375
Handheld Communicator, in AMS, or in the Rosemount Radar Master (RRM)
program, are shown in Table 7-3. Errors normally result in Analog Output
alarm.
Errors are indicated in RRM in the
Diagnostics
window.
Table 7-3. Error messages.
Message
Description
Action
RAM Error
An error in the gauge data memory
(RAM) has been detected during the
startup tests. Note: this resets the
gauge automatically.
Contact Emerson Process
Management Service Department.
FPROM Error
An error in the gauge program
memory (FPROM) has been detected
during the startup tests. Note: this
resets the gauge automatically.
Contact Emerson Process
Management Service Department.
HREG Error
An error in the transmitter
configuration memory (EEPROM) has
been detected. The error is either a
checksum error that can be solved by
loading the default database or a
hardware error.
NOTE: the default values are used
until the problem is solved.
Load default database and restart the
transmitter. Contact Emerson
Process Management Service
Department if the problem persists.
MWM Error
An error in the microwave module.
Contact Emerson Process
Management Service Department.
Display Error
An error in the LCD.
Contact Emerson Process
Management Service Department.
Modem Error
Modem hardware failure.
Contact Emerson Process
Management Service Department.
Analog Out Error
An error in the Analog Out Module.
Contact Emerson Process
Management Service Department.
Internal Temp Error
An error in the internal temperature
measurement.
Contact Emerson Process
Management Service Department.
Other HW Error
An unspecified hardware error has
been detected.
Contact Emerson Process
Management Service Department.
Meas Error
A serious measurement error has
been detected.
Contact Emerson Process
Management Service Department.
Config Error
At least one configuration parameter
is outside allowed range.
NOTE: the default values are used
until the problem is solved.
• Load the default database and
restart the transmitter (see “Reset
to Factory Settings” on page 7-19)
• Configure the transmitter or upload
a backup configuration file (see
• Contact Emerson Process
Management Service Department if
the problem persists
SW Error
An error has been detected in the
transmitter software.
Contact Emerson Process
Management Service Department.
Содержание 5401
Страница 2: ......
Страница 4: ......
Страница 10: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series TOC 6 ...
Страница 24: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 2 10 ...
Страница 80: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 5 26 ...
Страница 90: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 6 10 ...
Страница 160: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series B 16 ...
Страница 174: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series C 14 ...
Страница 177: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Index 3 Rosemount 5400 Series ...
Страница 178: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series Index 4 ...
Страница 179: ......