Reference Manual
00809-0100-4026, Rev FA
July 2009
7-15
Rosemount 5400 Series
Figure 7-10. The Log Registers
function can be used to verify
that the transmitter is working
properly.
To begin logging:
1. Click the
Browse
button, select a directory to store the log file, and type
a log file title.
2. Click the
Select Register
button and choose the register type to be
logged.
3. Select the desired registers to be logged. There are three options
available:
Standard
,
Service
, and
Custom
. Standard and Service refer to
pre-defined sets of registers. The Custom option allows the user to
choose the desired range of registers.
4. Enter the update rate. An update rate of 10 seconds means that the plot
will update every 10 seconds.
5. Click the
Start Log
button. The logging will proceed until the
Stop Log
button is clicked.
BACKING UP THE
TRANSMITTER
CONFIGURATION
Use this RRM option to make a backup copy of the configuration parameters
in the transmitter database. The backup file can be used to restore the
transmitter configuration. It can also be used for configuration of a transmitter
in a similar application. Parameters in the saved file can be uploaded directly
to the new device.
The backup function is available from the
Device
menu in RRM.
Start Log
Browse
Select Register
Update rate
Содержание 5401
Страница 2: ......
Страница 4: ......
Страница 10: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series TOC 6 ...
Страница 24: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 2 10 ...
Страница 80: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 5 26 ...
Страница 90: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series 6 10 ...
Страница 160: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series B 16 ...
Страница 174: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series C 14 ...
Страница 177: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Index 3 Rosemount 5400 Series ...
Страница 178: ...Reference Manual 00809 0100 4026 Rev FA July 2009 Rosemount 5400 Series Index 4 ...
Страница 179: ......