291
May 2016
Reference Manual
00809-0100-4530, Rev DD
I7, IG
Intrinsically safe and FISCO model:
Certificate: IECEx NEM 06.0001X
Ex ia IIC T4 Ga (-50 °C
T
a
+70 °C
(1)
)
Ex ia/ib IIC T4 Ga/Gb
(2)
Ex ta IIIC T 79 °C
(3)
Da (-50 °C < T
a
< +70 °C
)
4-20 mA/HART model:
U
i
=
30 Vdc, I
i
=
130 mA, P
i
=
1.0 W, C
i
= 7.26 nF,
L
i
=0
H
F
OUNDATION
Fieldbus model:
U
i
=
30 Vdc, I
i
=
300 mA, P
i
=
1.5 W, C
i
= 4.95 nF,
L
i
= 0
H
FISCO model: U
i
= 17.5 Vdc, I
i
=
380 mA, P
i
= 5.32 W,
C
i
= 4.95 nF, L
i
< 0
H
Installation drawing: 9240 030-938
Approval valid for HART, F
OUNDATION
Fieldbus, and
FISCO options.
N7
Type n:
Certificate: IECEx NEM 10.0005X
Ex nA ic IIC T4 Gc
Ex tc IIIC T79°C
-50°C < T
a
< +70°C
4-20 mA/HART model: 42.4 Vdc, 23 mA
F
OUNDATION
Fieldbus model: 32 Vdc, 21 mA
Ex ic IIC T4 Gc
Ex tc IIIC T79°C
-50°C < T
a
< +70°C
4-20 mA/HART model: U
i
=
42.4 V, I
i
=
23 mA,
P
i
=
1.0 W, C
i
= 7.25 nF, L
i
= 0
H
F
OUNDATION
Fieldbus model: U
i
=
32 V, I
i
=
21 mA,
P
i
=
0.7 W, C
i
= 4.95 nF, L
i
= 0
H
Approval valid for HART and F
OUNDATION
Fieldbus
options.
B.4.8 Other certifications
U1
Overfill
prevention
Certificate: Z-65.16-476
TÜV-tested and approved by DIBt for overfill
prevention according to the German WHG
regulations.
Approval valid for HART and F
OUNDATION
Fieldbus
options.
Suitability for intended use
Compliant with NAMUR NE 95, version 07.07.2006 “Basic
Principles of Homologation”.
Type approval certifications (Marine/shipboard
approvals)
SBS
American Bureau of Shipping (ABS)
SDN
Det Norske Veritas (DNV)
SLL
Lloyd’s
Register
SBV
Bureau Veritas
B.5 Combination approvals
KA
ATEX, FM, CSA Flameproof/Explosion-proof
KB
ATEX, FM, IECEx Flameproof/Explosion-proof
KC
ATEX, CSA, IECEx Flameproof/Explosion-proof
KD
FM, CSA, IECEx Flameproof/Explosion-proof
KE
ATEX, FM, CSA Intrinsic Safety
KF
ATEX, FM, IECEx Intrinsic Safety
KG
ATEX, CSA, IECEx Intrinsic Safety
KH
FM, CSA, IECEx Intrinsic Safety
KI
FISCO - ATEX, FM, CSA Intrinsic Safety
KJ
FISCO - ATEX, FM, IECEX Intrinsic Safety
KK
FISCO - ATEX, CSA, IECEX Intrinsic Safety
KL
FISCO - FM, CSA, IECEX Intrinsic Safety
B.6 Approval drawings
This section contains a FM Approvals System Control
Drawing and Canadian Standards Association and
IECEx/ATEX Installation Drawings. You must follow the
installation guidelines presented in order to maintain
certified ratings for installed transmitters.
This section contains the following drawings:
Rosemount
™
drawing 9240 030-936:
System control drawing for hazardous location
installation of intrinsically safe FM approved apparatus.
Rosemount drawing 9240 030-937:
Installation drawing for hazardous location installation of
CSA approved apparatus.
Rosemount drawing 9240 030-938:
Installation drawing for hazardous location installation of
ATEX and IECEx approved apparatus.
Rosemount drawing 9240 031-957: Installation drawing
for hazardous location installation of ATEX and IECEx
approved apparatus, type n.
1.
+60 °C with F
OUNDATION
Fieldbus or FISCO option.
2.
Valid for the FISCO option only.
3.
+69 °C with F
OUNDATION
Fieldbus or FISCO option.
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...