107
Reference Manual
00809-0100-4530, Rev DD
May 2016
Probe
6.
Verify that the Probe Type and Probe Length are correct.
HART command: [2, 1, 2].
F
OUNDATION
Fieldbus parameters:
TRANSDUCER 1100 > PROBE_TYPE
TRANSDUCER 1100 > PROBE_LENGTH
Check that the correct probe type is chosen. Normally, the probe type is
pre-configured at the factory, but if the current configuration does not match the
actual probe, choose the correct
Probe Type
from the list.
The Rosemount 5300 Series Transmitter automatically makes some initial
calibrations based on the chosen
Probe Type
. The following probe types are
available:
Rigid twin
Flexible twin
Coaxial, coaxial HP, coaxial HTHP
Rigid single, rigid single HTHP/HP/C, rigid single PTFE
Flexible single, flexible single HTHP/HP/C, flexible single PTFE
The
Probe Length
is the distance from the upper reference point to the end of the
probe, see
. If the probe is anchored to a weight, do not include the
height of the weight. The probe length needs to be changed if, for example, the
probe is shortened.
Содержание 5300 Series
Страница 2: ......
Страница 12: ...Reference Manual 00809 0100 4530 Rev DD Contents May 2016 xii Contents ...
Страница 94: ...80 Reference Manual 00809 0100 4530 Rev DD Electrical Installation May 2016 Electrical Installation ...
Страница 162: ...148 Reference Manual 00809 0100 4530 Rev DD Operation May 2016 Operation ...
Страница 214: ...200 Reference Manual 00809 0100 4530 Rev DD Service and Troubleshooting May 2016 Service and Troubleshooting ...
Страница 310: ...Product Certifications May 2016 Reference Manual 00809 0100 4530 Rev DD 296 Product Certifications ...
Страница 358: ...Level Transducer Block May 2016 Reference Manual 00809 0100 4530 Rev DD 344 Level Transducer Block ...
Страница 431: ...5 Reference Manual 00809 0100 4530 Rev DD May 2016 ...