Reference Manual
00809-0100-4148, Rev BA
July 2010
Rosemount 148
B-4
NG
IECEx Type n Component
Certificate number: IECEx BAS08.0012U
Ex nA IIC T5 (-60 °C
T
amb
80 °C)
Ex nA IIC T6 (-60 °C
T
amb
60 °C)
Input Parameter: Ui = 42.4 Vdc
Schedule of Limitations:
The component must be housed in a suitably certified enclosure that provides a degree of
protection of at least IP54.
INSTALLATION
DRAWINGS
The installation guidelines presented by the drawings must be followed in
order to maintain certified ratings for installed transmitters.
Rosemount Drawing 00148-1055, Rev AA, 2 Sheets
Factory Mutual Intrinsic Safety and Non-Incendive Installation Drawing
Rosemount Drawing 00148-1065, Rev AA, 1 Sheet
Factory Mutual Explosion-Proof Installation Drawing
Rosemount Drawing 00148-1056, Rev AA, 1 Sheet
CSA Explosion-Proof and Non-Incendive Installation Drawing
Rosemount Drawing 00148-1057, Rev AA, 1 Sheet
IECEX Intrinsic Safety Installation Drawing
Rosemount Drawing 00644-1059, Rev AH, 1 Sheet
CSA Explosion-Proof Installation Drawing
IMPORTANT
Once a device labeled with multiple approval types is installed, it should not
be reinstalled using any of the other labeled approval types. To ensure this,
the approval label should be permanently marked to distinguish the used from
the unused approval type(s).
Содержание 148
Страница 2: ......
Страница 4: ......
Страница 12: ...Reference Manual 00809 0100 4148 Rev BA July 2010 Rosemount 148 1 6 ...
Страница 28: ...Reference Manual 00809 0100 4148 Rev BA July 2010 Rosemount 148 3 6 ...
Страница 40: ...Reference Manual 00809 0100 4148 Rev BA July 2010 Rosemount 148 A 10 ...
Страница 51: ......