MODEL SOLU COMP II
SECTION 8.0
TROUBLESHOOTING
SECTION 8.0
TROUBLESHOOTING
8.1 OVERVIEW
8.2 TROUBLESHOOTING USING FAULT CODES
8.3 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — DO
8.4 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — pH
8.5 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — GENERAL
8.6 SIMULATING INPUTS — DISSOLVED OXYGEN
8.7 SIMULATING INPUTS — pH
8.8 SIMULATING TEMPERATURE
8.9 MEASURING REFERENCE VOLTAGE — pH
8.1 OVERVIEW
The Solu Comp II continuously monitors itself and the sensor for faults. When the analyzer detects a fault, the
word
fault appears in the display alternately with the measurement. If alarm 3 was configured as a fault alarm,
the alarm relay will energize. The outputs do not change during a fault condition. They continue to reflect the
measured dissolved oxygen, pH, ORP, or temperature. Press
"
"
to display the fault codes.
8.2 TROUBLESHOOTING USING FAULT CODES
8.2.1 Sensor Current Exceeds 210
µµ
A
Excessive sensor current implies that the sensor is miswired or the sensor has failed.
8.2.2 Absolute Value of Measured Voltage Exceeds 2500 mV
The voltage of a pH cell is usually between 600 mV and -600 mV, and the voltage of an ORP cell is between 2000
and -2000 mV. Readings outside the range -2500 mV to 2500 mV usually indicate a problem with sensor wiring or
analyzer electronics.
A. If the sensor is being installed for the first time, check the wiring connections. See Section 3.3. If a junction box
is being used, check connections at the junction box, too.
B. If the preamplifier is in a junction box, verify that the wires supplying the power (±5 Vdc) to the preamplifier are
connected at the analyzer and junction box.
C. Verify that the sensor is completely submerged in the process liquid.
Fault Code
Explanation
See Section
S1 Out of Range
Sensor current exceeds 210
µ
A (dissolved oxygen only)
8.2.1
S2 Out of Range
Absolute value of measured voltage exceeds 2500 mV (pH/ORP only)
8.2.2
S2 Broken Glass
pH sensitive glass membrane is broken
8.2.3
TC1 Open or TC2 Open
RTD for sensor 1 or sensor 2 is open
8.2.4
TC1 Shorted or TC2 Shorted
RTD for sensor 1 or sensor 2 is shorted
8.2.4
S1 or S2 Sense Line Open
RTD sense line for sensor 1 or sensor 2 is open
8.2.5
EEPROM Failure
EEPROM failure
8.2.6