Steps 1 – 4 should be performed with the
device turned OFF. Once the unit has been
turned ON, do not remove electrodes
without first turning OFF the device.
IMPORTANT!
QUICK REFERENCE GUIDE
PAGE 2
2
CONNECT ELECTRODES
TO LEAD WIRES
Firmly insert
the end of
the lead wire
pin into the
electrode
connectors.
Make sure no bare metal of the pins are exposed.
3
CONNECT LEAD WIRES
TO TOP OF DEVICE
4
PLACE ELECTRODES
ON SKIN
6
SELECT THE
THERAPEUTIC MODE
Remove electrodes from
protective backing. Keep
backing to properly store
electrodes after use.
Apply electrodes to the
exact site indicated by
your doctor or therapist.
Skin at the application site
should be clean and dry.
There are 2 therapeutic modes available: TENS and
MIC (microcurrent). Press the
M
button and continue
to press until you have advanced to your desired
therapeutic mode.
Consult your physician for your suitable
therapeutic mode.
Press the
S
button to enter and cycle through
to each setting state.
SET THERAPEUTIC PROGRAM
There are 4 programs available in the TENS therapeutic mode:
B
-
Burst,
N
-
Normal,
M
-
Pulse Width Modulation, and
M1
-
Pulse Rate Modulation. The therapeutic
program can be selected by pressing the
s
and
t
buttons.
SET CYCLE TIME (OPTIONAL)
Cycle time is adjustable from
5
to
30
seconds. Only modulation has this parameter
setting. Press
S
button to enter this menu, and then press the
s
and
t
buttons to
adjusting the setting.
SET TIMER
Press the
S
button to enter this setting. The treatment time is adjustable from
1
to
60
minutes or Continuous. Press the
s
and
t
buttons to adjust setting. You
can set the timer to “Continuous” mode by pressing the
s
control when it shows
60 minutes. The output will be shut off when time is up. If in continuous mode, you
need to manually turn off device.
SET PULSE WIDTH
Pulse Width is adjustable from
50
uS to
300
uS. Press
S
button to enter this menu,
then press the
s
and
t
buttons to adjust the setting.
5
TURN DEVICE ON
This unit operates with
one or two sets of
electrodes. When using
one set, plug lead wire
into Channel 1 on top of
the unit. Use Channel 2
to operate a second set.
Both channels may be
used to stimulate two
different areas at the
same time.
TOP VIEW
See
Electrode Placement Guide
for more information.
1
INSERT BATTERY
Press arrow and slide battery compartment
cover open. Insert 9V battery, matching the
+ and – ends to those in unit.
TENS SETTING
2
1
3
4
SET PULSE RATE
Pulse rate is adjustable from
1
Hz to
150
Hz (0.5 Hz to 5 Hz for Burst). Press the
S
button after scrolling through and completing all settings to start the therapy
program. The
s
and
t
buttons can be used to increase
s
or decrease
t
the
intensity level.
5
1
3
4
3
5
Check that electrodes are securely placed on
the skin before activating device. Press and release the
button in the center of the unit to turn it ON.