Page 25
© ROPOX 2017
13.
Cleaning/ maintenance
13.1 Cleaning of frame
Wipe the frame with a wrung cloth with lukewarm water and a mild detergent. Do not use
enchants or abrasives or grinding cloths, brushes or sponges.
After cleaning, dry the frame.
13.2 Maintenance
The frame is maintenance free. Moving parts have been lubricated for life. For reasons of safety
and reliability, we recommend inspection of the frame once a year:
➢
Check that all bolts have been securely fastened.
➢
Check that the frame moves freely and without problems from bottom to top position.
➢
Resetting the frame:
Press the UP button and adjust the frame to top position. Press
the UP button once more and keep it depressed for 5 sec. to reset the control unit.
Check that the movement is smooth and even. Make sure that the mains cable moves
freely.
➢
Check that all cables have been mounted correctly and are undamaged.
After each inspection, complete the service schedule.
Only use Ropox original spare parts for replacement. If other spare parts are used, the right to
complain may become void.
WARNING
Mains voltage must NOT be connected during cleaning and electrical
components must NOT be flushed with water
Inspections, service and repairs must be carried out by competent personnel
Make sure that the load on the unit
is not uneven. The load must be
evenly distributed.
=
=
=