Ropox AdaptLine 40-42010-2 Скачать руководство пользователя страница 6

TF 200.01.0042_eng 

3.  General requirements 

 

3.1 Product information 

Manufacturer 

Ropox A/S  
Ringstedgade 221, DK-4700 Naestved 
+45 55 75 05 00 
[email protected] 

Product models 

Part number 

Model 

Configuration 

40-42010-2 
40-42011-2 
40-42012-2 
40-42110-2 

Ropox AdaptLine Vask Complete 
Ropox AdaptLine Vask Complete 
Ropox AdaptLine Vask Complete 
Ropox AdaptLine Unit 

Standard 
Support 
Hospital 

Expected service life 

5 years 

MDR Class 
2017/745 

Class I 

Intended environments 

This product is to be used only in: 
 
Professional Healthcare Environment 
Homecare Environment 
 

Maximum user weight 
according to 
EN17966:2016 

40-42010-2 = 165 kg / 363 lbs 
40-42011-2 = 165 kg / 363 lbs 
40-42012-2 = 120 kg / 264 lbs 
40-42110-2 = 200 kg / 440 lbs 

Height adjustment 

20 cm / 7,8 in 

Ambient temperature 
range 

-5

 °C to +4

0 °C Transport and storage 

 

+5

 °C to +

40 

°

C Operation 

 

Relative humidity: 

20% to 80% - non-condensing 

Materials in contact 
with patient 

Wash basin 
 
 
 
Mixer tap 
 
Cover 
 
Top cover 

40-42010: Porcelain 
40-42011: Synthetic/marble composite, with gel coat surface 
40-42012: Porcelain 
 
Chrome 
 
St37 tube construction with powder coating 
 
ABS 

Accessories 

Flexible handle 
 
Plumbing kit 

40-42170 
 
40-25731 

 

3.2 Product description 

AdaptLine is a height-adjustable unit where you can mount a disability-friendly washbasin. AdaptLine has a 
convenient and compact design which makes it highly suitable for smaller bathrooms where space-saving 
solutions are required. 
 
 

3.3 Intended use 

The AdaptLine washbasins are available with manual height adjustment to better the use for wheelchair users 
and walking impaired patient. Height adjustment is done with the small crank. The height adjustment range is 
20 cm / 7,8 in. 
 
 

3.4 Intended operator 

Reading and understanding this manual will be sufficient training for an intended operator. Children or patient 
are not considered intended operator. 

 

 

Содержание AdaptLine 40-42010-2

Страница 1: ...TF 200 01 0042_eng Dato 29 07 2022 Ropox AdaptLine Manual User Manual and Mounting Instructions This manual must always accompany the product...

Страница 2: ...3 INTENDED USE 6 3 4 INTENDED OPERATOR 6 3 5 DIMENSION DRAWINGS 7 3 6 USE 9 3 7 SAFETY IN USE 10 3 8 CLEANING 10 4 MOUNTING INSTRUCTIONS 11 4 1 INSTALLATION INFORMATION 11 4 2 CONNECTION ADAPTERS 12 4...

Страница 3: ...rsons installing and or using this product either as operator or user should have the necessary safety information and access to this manual If the supply cord is damaged it must be replaced by the ma...

Страница 4: ...nce Cleaning and user maintenance shall not be made by children without supervision Do not use the product if it has defects or have become damaged before being repaired or replaced If the control box...

Страница 5: ...porated in Directive 2007 47 EEC European Medical Device Regulation EU 2017 745 Council Directive 2006 42 EC on machinery Council Directive 2011 65 EU RoHS Manufacturer name and address Date of produc...

Страница 6: ...C Operation Relative humidity 20 to 80 non condensing Materials in contact with patient Wash basin Mixer tap Cover Top cover 40 42010 Porcelain 40 42011 Synthetic marble composite with gel coat surfa...

Страница 7: ...7 TF 200 01 0042_eng 3 5 Dimension drawings 40 42110 40 42010 NB All dimensions are in cm in 38 cm 15 in 28 cm 11 in 31 2 cm 12 3 in 5 cm 2 in 35 cm 13 8 in 1 2 cm 0 5in...

Страница 8: ...8 TF 200 01 0042_eng 40 42011 40 42012 NB All dimensions are in cm in...

Страница 9: ...ismounted first however Then mount the flexible handle on the hexagonal shaft Subsequently the washbasin height is adjusted by turning the handle Note Be careful not to swing the crank into the mirror...

Страница 10: ...may be wiped with a wrung cloth with lukewarm water and a mild detergent Do not use corrosive grinding fluids or scouring sponges Disinfection may be done with household disinfectants Cleaning of was...

Страница 11: ...ain is 32 mm 1 3 in The water supply ends with 2 two 90 stop valves with outside thread pointing downward All hoses for water and drain installations must be flexible to ensure free and unhindered adj...

Страница 12: ...from washbasin Coupling piece Extra hose 1 2 1 2 Item no 97001669 1 2 x 1 2 05m steel Item no 97001123 1 2 x 1 2 1m plastic Item no 97001120 Drain adapters Drain from washbasin Adapter Installation in...

Страница 13: ...0 A 40 42011 B 40 42012 C 1 one mixer tap Oras Saga 3910F 1 one plumbing kit 1 one 97001601 Siphon 1 one 97001162 Flexible drain hose 32 2 two 97001035 Drain elbow 88 5 2 two 97001012 Nipple to 3 8 2...

Страница 14: ...ing instructions Prior to the installation please make sure that all parts have been provided in accordance with the list of components 300 mm 11 8 in 69 mm 2 7 in 1000 mm 39 4 in 908 mm 35 7 in 610 m...

Страница 15: ...15 TF 200 01 0042_eng Caution When mounting a porcelain washbasin an insolation foam pad MUST be placed between the washbasin and the unit It is recommended to grout around the washbasin...

Страница 16: ...16 TF 200 01 0042_eng 500 mm 19 7 in 250 mm 9 8 in 180 mm 7 1 in 90 mm 3 5 in...

Страница 17: ...0 42010 2 40 42011 2 and 40 42012 2 Caution Inspections service and repairs must be carried out by qualified personnel Description In product Item no 1 Ropox AdaptLine Unit Manual 1 40 42110 1 2 Mixer...

Страница 18: ...disposal of components must be done by waste disposal professionals Part Disposal recommendation Wash basin Ceramic Unit Steel waste Top cover Plastic 8 Complaints See our general terms of sale and d...