Electromagnetic compatibility
10.1 Suitable Environments
The device is suitable for use at home, at daycare centers, at day centers for persons with physical or mental
disabilities or at hospitals except near active HF surgical equipment and the RF shielded room of a medical
electrical system for magnetic resonance imaging, where the intensity of electromagnetic disturbances is
high.
10.2 Basic safety and Essential performance
The device does not have any function related to basic safety or essential performance. In case of loss or
degradation of performance, the user can get on and off the product.
10.3 Adjacent and stacked use
10.4 Cables
Cable component
Length
Power cable
3,2m
Hand control spiral cable
Approx. 5m extended.
Warning!
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
10.5 RF portable equipment
Warning !
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of
the device, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
10.6 Compliance for emission and immunity standard
The product is tested and found compliant for EMC according to EN60601-1-2:2015 for Clause 7 and Clause
8. Documentation is tested and found compliant according to EN60601-1-2:2015 for Clause 5
10.6.1 Emission class and group
Phenomenon
Test method
Class
Group
Conducted RF emissions EN 55011
Class B
Group 1
Radiated RF emissions
EN 55011
Class B
Group 1
Harmonic current
emissions
EN 61000-3-2
Class A
-
Voltage changes, voltage
fluctuations and flicker
emissions
EN 61000-3-3
-
-
Warning !
Use of this equipment adjacent to or stacked with other equipment should be voided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.