Chapter 2 Accura 2350-IDC Operation/Setup
Accura 2350-IDC User Guide
Page 14
ⓒ
2014 Rootech Inc. All Rights Reserved
Item
Description
Connector type
Terminal block
Wire specification
1.5 to 3.5 mm
2
(16 to 12 AWG), copper or aluminum
Input current
DC voltage -50 V to 50 V
Isolation/channel
AC 2,000 V for 1 minute, 8 channels
(differential inputs, non-isolated between channels)
Input resistance
Higher than 2 M
Ω
Accuracy
±
0.5% FS
1
Event
“
Over
”
/
”
Under
”
2
event selectable based on the threshold value for each channel
1. Full Scale (FS) indicates 20 mA.
2. An
“
Over
”
event indicates an event where the measured value rises above the threshold. An
“
Under
”
event
indicates an event where the measured value falls below the threshold. Over and Under current events consist of start
and end events. Either event can be configured.
Fig 2.2 Analog Input DC Current
Meter
DC Current
-20 mA ~ 20 mA
Measurement
Input resistance
150
Ω
I
DC1
I
DC1G